Accuracy and Safety of Dexcom G7 Continuous Glucose Monitoring in Adults with Diabetes

Abstract

We evaluated the accuracy and safety of a seventh generation (G7) Dexcom continuous glucose monitor (CGM) during 10.5 days of use in adults with diabetes. Adults with either type 1 or type 2 diabetes (on intensive insulin therapy or not) participated at 12 investigational sites in the United States. In-clinic visits were conducted on days 1 or 2, 4 or 7, and on the second half of day 10 or the first half of day 11 for frequent comparisons with comparator blood glucose measurements obtained with the YSI 2300 Stat Plus glucose analyzer. Participants wore sensors concurrently on the upper arm and abdomen. Accuracy evaluation included the proportion of CGM values within 15% of comparator glucose levels >100 mg/dL or within 15 mg/dL of comparator levels ≤100 mg/dL (%15/15), along with the %20/20 and %30/30 agreement rates. The mean absolute relative difference (MARD) between temporally matched CGM and comparator values was also calculated. Data from 316 participants (619 sensors, 77,774 matched pairs) were analyzed. For arm- and abdomen-placed sensors, overall MARDs were 8.2% and 9.1%, respectively. Overall %15/15, %20/20, and %30/30 agreement rates were 89.6%, 95.3%, and 98.8% for arm-placed sensors and were 85.5%, 93.2%, and 98.1% for abdomen-placed sensors. Across days of wear, glucose concentration ranges, and rates of change, %20/20 agreement rates varied by no more than 9% from the overall %20/20. No serious adverse events were reported. The G7 CGM provides accurate glucose readings with single-digit MARD with arm or abdomen placement in adults with diabetes. Clinicaltrials.gov: NCT04794478.

Citing Articles

Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes.

Eichenlaub M, Waldenmaier D, Wehrstedt S, Pleus S, Link M, Jendrike N J Diabetes Sci Technol. 2025; :19322968251315459.

PMID: 39902649 PMC: 11795573. DOI: 10.1177/19322968251315459.


International Society for Pediatric and Adolescent Diabetes Clinical Practice Consensus Guidelines 2024 Diabetes Technologies: Glucose Monitoring.

Tauschman M, Cardona-Hernandez R, DeSalvo D, Hood K, Laptev D, Lindholm Olinder A Horm Res Paediatr. 2025; 97(6):615-635.

PMID: 39884260 PMC: 11854985. DOI: 10.1159/000543156.


Healthy lifestyle before and during pregnancy to prevent childhood obesity: study protocol for a parallel group randomised trial - the PRE-STORK trial.

Kornerup N, Danielsen J, Sahl R, Pico M, Johansen M, Knop F BMJ Open. 2025; 15(1):e087895.

PMID: 39863406 PMC: 11784338. DOI: 10.1136/bmjopen-2024-087895.


Experiences and opinions of adults with type 1 diabetes on the android-based open-source closed-loop system in China: a qualitative study.

Huang Q, Ni Y, Lei M, Ling P, Yan J, Guo X BMJ Open. 2025; 15(1):e094333.

PMID: 39819904 PMC: 11751815. DOI: 10.1136/bmjopen-2024-094333.


International Society for Pediatric and Adolescent Diabetes Clinical Practice Consensus Guidelines 2024: Glycemic Targets.

de Bock M, Agwu J, Deabreu M, Dovc K, Maahs D, Marcovecchio M Horm Res Paediatr. 2024; 97(6):546-554.

PMID: 39701064 PMC: 11854972. DOI: 10.1159/000543266.


References
1.
Martens T, Beck R, Bailey R, Ruedy K, Calhoun P, Peters A . Effect of Continuous Glucose Monitoring on Glycemic Control in Patients With Type 2 Diabetes Treated With Basal Insulin: A Randomized Clinical Trial. JAMA. 2021; 325(22):2262-2272. PMC: 8173473. DOI: 10.1001/jama.2021.7444. View

2.
Castorino K, Polsky S, OMalley G, Levister C, Nelson K, Farfan C . Performance of the Dexcom G6 Continuous Glucose Monitoring System in Pregnant Women with Diabetes. Diabetes Technol Ther. 2020; 22(12):943-947. PMC: 7757524. DOI: 10.1089/dia.2020.0085. View

3.
Litchman M, Wawrzynski S, Woodruff W, Arrington J, Nguyen Q, Gee P . Continuous Glucose Monitoring in the Real World Using Photosurveillance of #Dexcom on Instagram: Exploratory Mixed Methods Study. JMIR Public Health Surveill. 2019; 5(2):e11024. PMC: 6555117. DOI: 10.2196/11024. View

4.
Wadwa R, Laffel L, Shah V, Garg S . Accuracy of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System During 10 Days of Use in Youth and Adults with Diabetes. Diabetes Technol Ther. 2018; 20(6):395-402. PMC: 6110124. DOI: 10.1089/dia.2018.0150. View

5.
Steineck I, Mahmoudi Z, Ranjan A, Schmidt S, Jorgensen J, Norgaard K . Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites. Diabetes Technol Ther. 2019; 21(5):295-302. DOI: 10.1089/dia.2019.0014. View