Prediction of Motion Induced Magnetic Fields for Human Brain MRI at 3 T

Overview
Journal MAGMA
Publisher Springer
Date 2023 Mar 25
PMID 36964797
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Maps of B field inhomogeneities are often used to improve MRI image quality, even in a retrospective fashion. These field inhomogeneities depend on the exact head position within the static field but acquiring field maps (FM) at every position is time consuming. Here we propose a forward simulation strategy to obtain B predictions at different head-positions.

Methods: FM were predicted by combining (1) a multi-class tissue model for estimation of tissue-induced fields, (2) a linear k-space model for capturing gradient imperfections, (3) a dipole estimation for quantifying lower-body perturbing fields (4) and a position-dependent tissue mask to model FM alterations caused by large motion effects. The performance of the combined simulation strategy was compared with an approach based on a rigid body transformation of the FM measured in the reference position to the new position.

Results: The transformed FM provided inconsistent results for large head movements (> 5° rotation, approximately), while the simulation strategy had a superior prediction accuracy for all positions. The simulated FM was used to optimize B shims with up to 22.2% improvement with respect to the transformed FM approach.

Conclusion: The proposed simulation strategy is able to predict movement-induced B field inhomogeneities yielding more precise estimates of the ground truth field homogeneity than the transformed FM.

References
1.
Jezzard P, Clare S . Sources of distortion in functional MRI data. Hum Brain Mapp. 1999; 8(2-3):80-5. PMC: 6873315. DOI: 10.1002/(sici)1097-0193(1999)8:2/3<80::aid-hbm2>3.0.co;2-c. View

2.
Glover G . 3D z-shim method for reduction of susceptibility effects in BOLD fMRI. Magn Reson Med. 1999; 42(2):290-9. DOI: 10.1002/(sici)1522-2594(199908)42:2<290::aid-mrm11>3.0.co;2-n. View

3.
Stenger V, Boada F, Noll D . Three-dimensional tailored RF pulses for the reduction of susceptibility artifacts in T(*)(2)-weighted functional MRI. Magn Reson Med. 2000; 44(4):525-31. PMC: 3044439. DOI: 10.1002/1522-2594(200010)44:4<525::aid-mrm5>3.0.co;2-l. View

4.
Jezzard P, Balaban R . Correction for geometric distortion in echo planar images from B0 field variations. Magn Reson Med. 1995; 34(1):65-73. DOI: 10.1002/mrm.1910340111. View

5.
Stockmann J, Wald L . In vivo B field shimming methods for MRI at 7T. Neuroimage. 2017; 168:71-87. PMC: 5760477. DOI: 10.1016/j.neuroimage.2017.06.013. View