Abstract

Per- and polyfluoroalkyl substances (PFAS) are persistent environmental pollutants associated with negative health impacts. Assessments of tubing-related measurement bias for volatile PFAS are lacking, as gas-wall interactions with tubing can delay quantification of gas-phase analytes. We use online iodide chemical ionization mass spectrometry measurements to characterize tubing delays for three gas-phase oxygenated PFAS - 4:2 fluorotelomer alcohol (4:2 FTOH), perfluorobutanoic acid (PFBA), and hexafluoropropylene oxide dimer acid (HFPO-DA). Perfluoroalkoxy alkane and high-density polyethylene tubing yielded relatively short absorptive measurement delays, with no clear dependence on tubing temperature or sampled humidity. Sampling through stainless steel tubing led to prolonged measurement delays due to reversible adsorption of PFAS to the tubing surface, with strong dependence on tubing temperature and sample humidification. Silcosteel tubing afforded shorter measurement delays than stainless steel due to diminished surface adsorption of PFAS. Characterizing and mitigating these tubing delays is crucial for reliable quantification of airborne PFAS. Per- and polyfluoroalkyl substances (PFAS) are persistent environmental contaminants. Many PFAS are sufficiently volatile to exist as airborne pollutants. Measurements and quantification of airborne PFAS can be biased from material-dependent gas-wall interactions with sampling inlet tubing. Thus, characterizing these gas-wall interactions are crucial for reliably investigating emissions, environmental transport, and fates of airborne PFAS.

Citing Articles

Measuring short-chain per- and polyfluoroalkyl substances in Central New Jersey air using chemical ionization mass spectrometry.

Mattila J, Offenberg J J Air Waste Manag Assoc. 2024; 74(8):531-539.

PMID: 38905230 PMC: 11412083. DOI: 10.1080/10962247.2024.2366491.


A review of sample collection and analytical methods for detecting per- and polyfluoroalkyl substances in indoor and outdoor air.

Ariel Geer Wallace M, Smeltz M, Mattila J, Liberatore H, Jackson S, Shields E Chemosphere. 2024; 358:142129.

PMID: 38679180 PMC: 11513671. DOI: 10.1016/j.chemosphere.2024.142129.


Characterizing Volatile Emissions and Combustion Byproducts from Aqueous Film-Forming Foams Using Online Chemical Ionization Mass Spectrometry.

Mattila J, Krug J, Roberson W, Burnette R, McDonald S, Virtaranta L Environ Sci Technol. 2024; 58(8):3942-3952.

PMID: 38350647 PMC: 10985785. DOI: 10.1021/acs.est.3c09255.

References
1.
Davis K, Aucoin M, Larsen B, Kaiser M, Hartten A . Transport of ammonium perfluorooctanoate in environmental media near a fluoropolymer manufacturing facility. Chemosphere. 2007; 67(10):2011-9. DOI: 10.1016/j.chemosphere.2006.11.049. View

2.
Lesmeister L, Lange F, Breuer J, Biegel-Engler A, Giese E, Scheurer M . Extending the knowledge about PFAS bioaccumulation factors for agricultural plants - A review. Sci Total Environ. 2020; 766:142640. DOI: 10.1016/j.scitotenv.2020.142640. View

3.
Gluge J, Scheringer M, Cousins I, DeWitt J, Goldenman G, Herzke D . An overview of the uses of per- and polyfluoroalkyl substances (PFAS). Environ Sci Process Impacts. 2020; 22(12):2345-2373. PMC: 7784712. DOI: 10.1039/d0em00291g. View

4.
Shoeib M, Harner T, Wilford B, Jones K, Zhu J . Perfluorinated sulfonamides in indoor and outdoor air and indoor dust: occurrence, partitioning, and human exposure. Environ Sci Technol. 2005; 39(17):6599-606. DOI: 10.1021/es048340y. View

5.
Riedel T, Lang J, Strynar M, Lindstrom A, Offenberg J . Gas-Phase Detection of Fluorotelomer Alcohols and Other Oxygenated Per- and Polyfluoroalkyl Substances by Chemical Ionization Mass Spectrometry. Environ Sci Technol Lett. 2019; 6(5):289-293. PMC: 6550326. DOI: 10.1021/acs.estlett.9b00196. View