Clinical Validation of an Artificial Intelligence-assisted Algorithm for Automated Quantification of Left Ventricular Ejection Fraction in Real Time by a Novel Handheld Ultrasound Device

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: We sought to evaluate the reliability and diagnostic accuracy of a novel handheld ultrasound device (HUD) with artificial intelligence (AI) assisted algorithm to automatically calculate ejection fraction (autoEF) in a real-world patient population.

Methods And Results: We studied 100 consecutive patients (57 ± 15 years old, 61% male), including 38 with abnormal left ventricular (LV) function [LV ejection fraction (LVEF) < 50%]. The autoEF results acquired using the HUD were independently compared with manually traced biplane Simpson's rule measurements on cart-based systems to assess method agreement using intra-class correlation coefficient (ICC), linear regression analysis, and Bland-Altman analysis. The diagnostic accuracy for the detection of LVEF <50% was also calculated. Test-retest reliability of measured EF by the HUD was assessed by calculating the ICC and the minimal detectable change (MDC). The ICC, linear regression analysis, and Bland-Altman analysis revealed good agreement between autoEF and reference manual EF (ICC = 0.85;  = 0.87,  < 0.001; mean bias -1.42% with limits of agreement 14.5%, respectively). Detection of abnormal LV function (EF < 50%) by autoEF algorithm was feasible with sensitivity 90% (95% CI 75-97%), specificity 87% (95% CI 76-94%), PPV 81% (95% CI 66-91%), NPV 93% (95% CI 83-98%), and a total diagnostic accuracy of 88%. Test-retest reliability was excellent (ICC = 0.91,  < 0.001;  = 0.91,  < 0.001; mean difference ± SD: 0.54% ± 5.27%,  = 0.308) and MDC for LVEF measurement by autoEF was calculated at 4.38%.

Conclusion: Use of a novel HUD with AI-enabled capabilities provided similar LVEF results with those derived by manual biplane Simpson's method on cart-based systems and shows clinical potential.

Citing Articles

Artificial Intelligence in Clinics: Enhancing Cardiology Practice.

Sakamoto A, Nakamura Y, Sato E, Kagiyama N JMA J. 2025; 8(1):131-140.

PMID: 39926098 PMC: 11799705. DOI: 10.31662/jmaj.2024-0190.


Assessment of an Artificial Intelligence Tool for Estimating Left Ventricular Ejection Fraction in Echocardiograms from Apical and Parasternal Long-Axis Views.

Vega R, Kwok C, Hareendranathan A, Nagdev A, Jaremko J Diagnostics (Basel). 2024; 14(16).

PMID: 39202209 PMC: 11353168. DOI: 10.3390/diagnostics14161719.


Multicenter validation study for automated left ventricular ejection fraction assessment using a handheld ultrasound with artificial intelligence.

Kagiyama N, Abe Y, Kusunose K, Kato N, Kaneko T, Murata A Sci Rep. 2024; 14(1):15359.

PMID: 38965290 PMC: 11224326. DOI: 10.1038/s41598-024-65557-5.


Artificial intelligence-assisted evaluation of cardiac function by oncology staff in chemotherapy patients.

Papadopoulou S, Dionysopoulos D, Mentesidou V, Loga K, Michalopoulou S, Koukoutzeli C Eur Heart J Digit Health. 2024; 5(3):278-287.

PMID: 38774364 PMC: 11104473. DOI: 10.1093/ehjdh/ztae017.


Experience with the Kosmos ultrasonographic tool in the approach and treatment of ambulatory patients of a heart failure clinic: a single-center cross-sectional study.

Gomez-Alvarez U, de la Fuente-Mancera J, Antonio-Villa N, Alvarez-Sangabriel A, Guizar-Sanchez C, Tenorio-Bautista F Arch Cardiol Mex. 2024; 94(1):79-85.

PMID: 38507324 PMC: 11160521. DOI: 10.24875/ACM.22000250.


References
1.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

2.
Knackstedt C, Bekkers S, Schummers G, Schreckenberg M, Muraru D, Badano L . Fully Automated Versus Standard Tracking of Left Ventricular Ejection Fraction and Longitudinal Strain: The FAST-EFs Multicenter Study. J Am Coll Cardiol. 2015; 66(13):1456-66. DOI: 10.1016/j.jacc.2015.07.052. View

3.
Otterstad J, Froeland G, St John Sutton M, Holme I . Accuracy and reproducibility of biplane two-dimensional echocardiographic measurements of left ventricular dimensions and function. Eur Heart J. 1997; 18(3):507-13. DOI: 10.1093/oxfordjournals.eurheartj.a015273. View

4.
Rahimi A, York M, Gheewala N, Markson L, Hauser T, Manning W . Trends in outpatient transthoracic echocardiography: impact of appropriateness criteria publication. Am J Med. 2011; 124(8):740-6. DOI: 10.1016/j.amjmed.2011.03.030. View

5.
Zannad F, Ferreira J, Pocock S, Anker S, Butler J, Filippatos G . SGLT2 inhibitors in patients with heart failure with reduced ejection fraction: a meta-analysis of the EMPEROR-Reduced and DAPA-HF trials. Lancet. 2020; 396(10254):819-829. DOI: 10.1016/S0140-6736(20)31824-9. View