» Articles » PMID: 23893844

Diagnostic Accuracy and Cost-effectiveness of a Pocket-sized Transthoracic Echocardiographic Imaging Device

Overview
Journal Clin Cardiol
Date 2013 Jul 30
PMID 23893844
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The recently introduced pocket-sized portable transthoracic echocardiography (pTTE) is accurate for measurement of cardiac chamber size and function as well as for assessment of valvular regurgitation. This study aimed to compare the diagnostic accuracy of the pocket-sized pTTE with the standard TTE (sTTE) and assess its cost-effectiveness.

Hypothesis: The use of pocket-sized pTTE, as an initial screening tool, may be feasible, accurate and cost-effective in the diagnostic strategy of cardiac abnormalities.

Methods: The study subjects were 200 patients scheduled for sTTE and an electrocardiogram (ECG). Each patient underwent pTTE examination with the Vscan (GE Medical Systems, Milwaukee, WI) immediately after sTTE. The findings of pTTE and the ECG were compared with the results of sTTE. Cost-effectiveness was calculated.

Results: There was a strong agreement in the detection of abnormal findings between pTTE and sTTE (agreement = 90%), whereas the agreement between the ECG and sTTE was 65%. When pTTE or the ECG was used as an initial screening tool prior to sTTE, similar cost reduction was obtained (approximately 30%) by reducing the number of referrals for sTTE. However, the negative predictive value of a diagnostic strategy with pTTE (92%) was superior to that with an ECG (67%).

Conclusions: This study demonstrates that the pocket-sized pTTE provides accurate detection of cardiac structural and functional abnormalities beyond the ECG. In addition, the use of pTTE as an initial screening tool prior to sTTE is cost-effective, suggesting that the pocket-sized pTTE is poised to alter the current diagnostic strategy in clinical practice.

Citing Articles

Clinical validation of an artificial intelligence-assisted algorithm for automated quantification of left ventricular ejection fraction in real time by a novel handheld ultrasound device.

Papadopoulou S, Sachpekidis V, Kantartzi V, Styliadis I, Nihoyannopoulos P Eur Heart J Digit Health. 2023; 3(1):29-37.

PMID: 36713988 PMC: 9707920. DOI: 10.1093/ehjdh/ztac001.


The Scope of Clinical Applications of Handheld Echocardiography.

Kim E J Cardiovasc Imaging. 2022; 30(1):35-36.

PMID: 35086167 PMC: 8792711. DOI: 10.4250/jcvi.2021.0113.


Handheld Echocardiography in a Clinical Practice Scenario: Concordances Compared to Standard Echocardiographic Reports.

Gavazzoni Blume G, Lechinewski L, Vieira I, Clausell N, Bertinato G, Machado-Junior P J Cardiovasc Imaging. 2022; 30(1):25-34.

PMID: 35086166 PMC: 8792719. DOI: 10.4250/jcvi.2020.0241.


Community-based aortic stenosis detection: clinical and echocardiographic screening during influenza vaccination.

Steeds R, Potter A, Mangat N, Frohlich M, Deutsch C, Bramlage P Open Heart. 2021; 8(1).

PMID: 34021069 PMC: 8144056. DOI: 10.1136/openhrt-2021-001640.


Screening for aortic stenosis using physical examination and echocardiography.

Abe Y J Echocardiogr. 2021; 19(2):80-85.

PMID: 33415574 DOI: 10.1007/s12574-020-00511-8.


References
1.
Greaves K, Jeetley P, Hickman M, Dwivedi G, Sabharwal N, Lim T . The use of hand-carried ultrasound in the hospital setting--a cost-effective analysis. J Am Soc Echocardiogr. 2005; 18(6):620-5. DOI: 10.1016/j.echo.2004.09.015. View

2.
Cardim N, Fernandez Golfin C, Ferreira D, Aubele A, Toste J, Cobos M . Usefulness of a new miniaturized echocardiographic system in outpatient cardiology consultations as an extension of physical examination. J Am Soc Echocardiogr. 2010; 24(2):117-24. DOI: 10.1016/j.echo.2010.09.017. View

3.
Andersen G, Haugen B, Graven T, Salvesen O, Mjolstad O, Dalen H . Feasibility and reliability of point-of-care pocket-sized echocardiography. Eur J Echocardiogr. 2011; 12(9):665-70. PMC: 3171198. DOI: 10.1093/ejechocard/jer108. View

4.
Tsutsui J, Maciel R, Costa J, Andrade J, Ramires J, Mathias Jr W . Hand-carried ultrasound performed at bedside in cardiology inpatient setting - a comparative study with comprehensive echocardiography. Cardiovasc Ultrasound. 2004; 2:24. PMC: 534795. DOI: 10.1186/1476-7120-2-24. View

5.
Fukuda S, Shimada K, Kawasaki T, Fujimoto H, Maeda K, Inanami H . Pocket-sized transthoracic echocardiography device for the measurement of cardiac chamber size and function. Circ J. 2009; 73(6):1092-6. DOI: 10.1253/circj.cj-08-1076. View