» Articles » PMID: 36610021

Characterization in Respect to Degradation of Titanium-coated Polypropylene Surgical Mesh Explanted from Humans

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Titanium-coated polypropylene (Ti-PP) mesh was introduced in 2002 as a surgical mesh for the treatment of hernias and shortly after for pelvic floor surgery, with the aim of improving biocompatibility when compared to non-titanised/regular PP mesh implants. The application of a titanium coating could also be beneficial to address concerns regarding the exposure of PP in an in vivo environment. Many studies have shown that PP, although it is widely accepted as a stable polymer, is subject to oxidation and degradation, such degradation affects the mechanical behavior, that is, the stiffness and tensile strength of PP mesh. Despite the wide clinical use of Ti-PP surgical meshes, no study has yet investigated the residual material properties post clinical deployment and subsequent explantation. In this study, two explanted Ti-PP mesh samples each having different incorporation durations from two patients were examined. Material analysis conducted within this study includes the following techniques: attenuated total reflectance-Fourier transform infrared spectroscopy (ATR-FTIR), Raman spectroscopy, low voltage - scanning electron microscopy (LV-SEM), backscattered electron (BSE) imaging, energy dispersive X-ray spectroscopy (EDS) and secondary election hyperspectral imaging (SEHI). The hypothesis of this study is that the Ti coating successfully shields the PP mesh from oxidative stress in vivo and thus protects it from degradation. The results of this analysis show for the first time evidence of bulk oxidation, surface degradation, and environmental stress cracking on explanted Ti-PP meshes.

Citing Articles

Strengthening preclinical testing to increase safety in surgical mesh.

Farr N, Workman V, Chapple C, MacNeil S, Rodenburg C Nat Rev Urol. 2024; 21(9):515-516.

PMID: 38693244 DOI: 10.1038/s41585-024-00889-5.


Assessing the Quality of Oxygen Plasma Focused Ion Beam (O-PFIB) Etching on Polypropylene Surfaces Using Secondary Electron Hyperspectral Imaging.

Farr N, Pasniewski M, de Marco A Polymers (Basel). 2023; 15(15).

PMID: 37571142 PMC: 10422415. DOI: 10.3390/polym15153247.


Development of PCL PolyHIPE Substrates for 3D Breast Cancer Cell Culture.

Jackson C, Ramos-Rodriguez D, Farr N, English W, Green N, Claeyssens F Bioengineering (Basel). 2023; 10(5).

PMID: 37237592 PMC: 10215449. DOI: 10.3390/bioengineering10050522.


Characterization in respect to degradation of titanium-coated polypropylene surgical mesh explanted from humans.

Farr N, Klosterhalfen B, Noe G J Biomed Mater Res B Appl Biomater. 2023; 111(5):1142-1152.

PMID: 36610021 PMC: 10952695. DOI: 10.1002/jbm.b.35221.

References
1.
Talley A, Rogers B, Iakovlev V, Dunn R, Guelcher S . Oxidation and degradation of polypropylene transvaginal mesh. J Biomater Sci Polym Ed. 2017; 28(5):444-458. DOI: 10.1080/09205063.2017.1279045. View

2.
Farr N, Klosterhalfen B, Noe G . Characterization in respect to degradation of titanium-coated polypropylene surgical mesh explanted from humans. J Biomed Mater Res B Appl Biomater. 2023; 111(5):1142-1152. PMC: 10952695. DOI: 10.1002/jbm.b.35221. View

3.
Dallenbach P . To mesh or not to mesh: a review of pelvic organ reconstructive surgery. Int J Womens Health. 2015; 7:331-43. PMC: 4386830. DOI: 10.2147/IJWH.S71236. View

4.
Kurtz J, Rael B, Lerma J, Wright C, Khraishi T, Auyang E . Effects of reactive oxygen species on the physical properties of polypropylene surgical mesh at various concentrations: a model for inflammatory reaction as a cause for mesh embrittlement and failure. Surg Endosc. 2015; 30(8):3250-5. DOI: 10.1007/s00464-015-4646-3. View

5.
Kaur M, Singh K . Review on titanium and titanium based alloys as biomaterials for orthopaedic applications. Mater Sci Eng C Mater Biol Appl. 2019; 102:844-862. DOI: 10.1016/j.msec.2019.04.064. View