Purpose: To evaluate the accuracy of axial length, keratometry, and refractive measurement with Myopia Master in ametropic children.

Methods: In this randomized prospective cross-sectional study, 125 children with ametropia (250 eyes) were recruited (55 boys and 70 girls; age range: 3-15 years). All examinations were performed under full cycloplegic conditions. Measurements of axial length (AL), keratometry, and autorefraction acquired with the Myopia Master were compared with those from the IOLMaster 500, IOLMaster 700, Nidek ARK-1, and manifest refraction. The differences between the different methods were analyzed, and their correlation was assessed by interclass correlation coefficients (ICCs), Bland-Altman plot, and correlation test.

Results: The ALs (mm) measured with Myopia Master, IOLMaster 500, and IOLMaster 700 were 23.67 ± 1.26, 23.68 ± 1.26, and 23.70 ± 1.25, respectively. The mean values and standard deviations for AL and keratometry readings from these devices were similar (P ≥ 0.059). The ICC analysis also revealed high consistency between the measurements (ICC ≥ 0.943). Additionally, the correlation coefficients were relatively high (r > 0.9, p < 0.001). Although the results of refraction obtained with the Myopia Master were slightly higher than those with manifest refraction (P ≤ 0.024), the agreement between these two measurements was excellent (ICC ≥ 0.858). The percentage of points outside the limits of agreements was < 5.22% in Bland-Altman plots for all analyses.

Conclusions: Myopia Master could be a highly efficient tool for clinical use as a three-in-one system (AL, keratometry, and refractive measurements) for screening in children with ametropia.

Citing Articles

Inter-Examiner and Inter-Instrument Agreement of the Myopia Master with the IOL Master 700 and NVisionK 5001 in Myopic Children.

McConnell E, Woods C, Doyle L, Fulton J, Saunders K, McCullough S Children (Basel). 2025; 12(2).

PMID: 40003223 PMC: 11854562. DOI: 10.3390/children12020121.


Comparative analysis of a novel spectral-domain OCT biometer versus swept-source OCT or OLCR biometer in healthy pediatric ocular biometry.

Long W, Ong A, Zheng B, Liang Z, Cui D Sci Rep. 2024; 14(1):25252.

PMID: 39448792 PMC: 11502762. DOI: 10.1038/s41598-024-77177-0.


Effects of Insufficient Sleep on Myopia in Children: A Systematic Review and Meta-Analysis.

Zhao X, He Y, Zhang J, Lin S, Zou H, Ma Y Nat Sci Sleep. 2024; 16:1387-1406.

PMID: 39308665 PMC: 11416795. DOI: 10.2147/NSS.S472748.


Comparison of Eye Axial Length Measurements Taken Using Partial Coherence Interferometry and OCT Biometry.

Rizzieri N, Facchin A Vision (Basel). 2024; 8(3).

PMID: 39189184 PMC: 11348256. DOI: 10.3390/vision8030046.


The effect of cycloplegia in the accuracy of autorefraction, keratometry and axial length using the Myopia Master.

Penaranda A, Torrado O, Marquez A, Baptista A, Serra P BMC Ophthalmol. 2024; 24(1):321.

PMID: 39090603 PMC: 11295713. DOI: 10.1186/s12886-024-03529-z.


References
1.
Dehnavi Z, Khabazkhoob M, Mirzajani A, Jabbarvand M, Yekta A, Jafarzadehpur E . Comparison of the Corneal Power Measurements with the TMS4-Topographer, Pentacam HR, IOL Master, and Javal Keratometer. Middle East Afr J Ophthalmol. 2015; 22(2):233-7. PMC: 4411623. DOI: 10.4103/0974-9233.151884. View

2.
Haigis W, Lege B, Miller N, Schneider B . Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000; 238(9):765-73. DOI: 10.1007/s004170000188. View

3.
Song J, Yoon D, Hyon J, Jeon H . Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900. Korean J Ophthalmol. 2020; 34(2):126-132. PMC: 7105785. DOI: 10.3341/kjo.2019.0102. View

4.
Gwiazda J, Hyman L, Hussein M, Everett D, Norton T, Kurtz D . A randomized clinical trial of progressive addition lenses versus single vision lenses on the progression of myopia in children. Invest Ophthalmol Vis Sci. 2003; 44(4):1492-500. DOI: 10.1167/iovs.02-0816. View

5.
Resnikoff S, Pascolini D, Mariotti S, Pokharel G . Global magnitude of visual impairment caused by uncorrected refractive errors in 2004. Bull World Health Organ. 2008; 86(1):63-70. PMC: 2647357. DOI: 10.2471/blt.07.041210. View