Accuracy of Heart Rate Measurement with Wrist-Worn Wearable Devices in Various Skin Tones: a Systematic Review

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Wearable consumer technology allows for the collection of a growing amount of personal health data. Through the analysis of reflected LED light on the skin, heart rate measurement and arrhythmia detection can be performed. Given that melanin alters skin light absorption, this study seeks to summarize the accuracy of cardiac data from wrist-worn wearable devices for participants of varying skin tones.

Methods: We conducted a systematic review, searching Embase, MEDLINE, CINAHL, and Cochrane for original studies that stratified heart rate and rhythm data for consumer wearable technology according to participant race and/or skin tone.

Results: A total of 10 studies involving 469 participants met inclusion criteria. The frequency-weighted Fitzpatrick score for skin tone was reported in six studies (n = 293), with a mean participant score of 3.5 (range 1-6). Overall, four of the ten studies reported a significant reduction in accuracy of heart rate measurement with wearable devices in darker-skinned individuals, compared to participants with lighter skin tones. Four studies noted no effect of user skin tone on accuracy. The remaining two studies showed mixed results.

Conclusions: Preliminary evidence is inconclusive, but some studies suggest that wearable devices may be less accurate for detecting heart rate in participants with darker skin tones. Higher quality evidence is necessary, with larger sample sizes and more objective stratification of participants by skin tone, in order to characterize potential racial bias in consumer devices.

Citing Articles

Atrial fibrillation burden in clinical practice, research, and technology development: a clinical consensus statement of the European Society of Cardiology Council on Stroke and the European Heart Rhythm Association.

Doehner W, Boriani G, Potpara T, Blomstrom-Lundqvist C, Passman R, Sposato L Europace. 2025; 27(3).

PMID: 40073206 PMC: 11901050. DOI: 10.1093/europace/euaf019.


Performance of a medical smartband with photoplethysmography technology and artificial intelligence algorithm to detect atrial fibrillation.

Blok S, Gielen W, Piek M, Hoeksema W, Tulevski I, Somsen G Mhealth. 2025; 11:5.

PMID: 39944860 PMC: 11811644. DOI: 10.21037/mhealth-24-10.


Validity of heart rate measurements in wrist-based monitors across skin tones during exercise.

Hung S, Serwa K, Rosenthal G, Eng J PLoS One. 2025; 20(2):e0318724.

PMID: 39928630 PMC: 11809914. DOI: 10.1371/journal.pone.0318724.


Issues and Limitations on the Road to Fair and Inclusive AI Solutions for Biomedical Challenges.

Faust O, Salvi M, Barua P, Chakraborty S, Molinari F, Rajendra Acharya U Sensors (Basel). 2025; 25(1.

PMID: 39796996 PMC: 11723364. DOI: 10.3390/s25010205.


Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis.

Singh S, Bennett M, Chen C, Shin S, Ghanbari H, Nelson B J Med Internet Res. 2024; 26:e62769.

PMID: 39388258 PMC: 11502980. DOI: 10.2196/62769.


References
1.
Khan Y, Han D, Pierre A, Ting J, Wang X, Lochner C . A flexible organic reflectance oximeter array. Proc Natl Acad Sci U S A. 2018; 115(47):E11015-E11024. PMC: 6255203. DOI: 10.1073/pnas.1813053115. View

2.
Sjoding M, Dickson R, Iwashyna T, Gay S, Valley T . Racial Bias in Pulse Oximetry Measurement. N Engl J Med. 2020; 383(25):2477-2478. PMC: 7808260. DOI: 10.1056/NEJMc2029240. View

3.
Fitzpatrick T . The validity and practicality of sun-reactive skin types I through VI. Arch Dermatol. 1988; 124(6):869-71. DOI: 10.1001/archderm.124.6.869. View

4.
Schaffarczyk M, Rogers B, Reer R, Gronwald T . Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women. Sensors (Basel). 2022; 22(17). PMC: 9459793. DOI: 10.3390/s22176536. View