Simple, Inexpensive, and Ergonomic Phantom for Quality Assurance Control of MRI Guided Focused Ultrasound Systems

Overview
Journal J Ultrasound
Publisher Springer
Specialty Radiology
Date 2022 Nov 4
PMID 36329304
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The popularity of Magnetic Resonance guided Focused Ultrasound (MRgFUS) as a beneficial therapeutic solution for many diseases is increasing rapidly, thus raising the need for reliable quality assurance (QA) phantoms for routine testing of MRgFUS systems. In this study, we propose a thin acrylic film as the cheapest and most easily accessible phantom for assessing the functionality of MRgFUS hardware and software.

Methods: Through the paper, specific QA tests are detailed in the framework of evaluating an MRgFUS preclinical robotic device comprising a single element spherically focused transducer with a nominal frequency of 2.75 MHz. These tests take advantage of the reflection of ultrasonic waves at a plastic-air interface, which results in almost immediate lesion formation on the film at a threshold of applied acoustic energy.

Results: The phantom offered qualitative information on the power field distribution of the FUS transducer and the ability to visualize different FUS protocols. It also enabled quick and reliable assessment of various navigation algorithms as they are used in real treatments, and also allowed for the assessment of the accuracy of robotic motion.

Conclusion: Therefore, it could serve as a useful tool for detecting defects in system's performance over its lifetime after establishing a baseline while concurrently contributing to establish QA and calibration guidelines for clinical routine controls.

References
1.
Ambrogio S, Baesso R, Bosio F, Fedele F, Ramnarine K, Zeqiri B . A standard test phantom for the performance assessment of magnetic resonance guided high intensity focused ultrasound (MRgHIFU) thermal therapy devices. Int J Hyperthermia. 2021; 39(1):57-68. DOI: 10.1080/02656736.2021.2017023. View

2.
Menikou G, Yiannakou M, Yiallouras C, Ioannides C, Damianou C . MRI-compatible bone phantom for evaluating ultrasonic thermal exposures. Ultrasonics. 2016; 71:12-19. DOI: 10.1016/j.ultras.2016.05.020. View

3.
Giannakou M, Drakos T, Menikou G, Evripidou N, Filippou A, Spanoudes K . Magnetic resonance image-guided focused ultrasound robotic system for transrectal prostate cancer therapy. Int J Med Robot. 2021; 17(3):e2237. PMC: 8244034. DOI: 10.1002/rcs.2237. View

4.
Menikou G, Yiannakou M, Yiallouras C, Ioannides C, Damianou C . MRI-compatible breast/rib phantom for evaluating ultrasonic thermal exposures. Int J Med Robot. 2017; 14(1). DOI: 10.1002/rcs.1849. View

5.
Antoniou A, Giannakou M, Evripidou N, Evripidou G, Spanoudes K, Menikou G . Robotic system for magnetic resonance guided focused ultrasound ablation of abdominal cancer. Int J Med Robot. 2021; 17(5):e2299. DOI: 10.1002/rcs.2299. View