Magnetic Resonance Image-guided Focused Ultrasound Robotic System for Transrectal Prostate Cancer Therapy

Overview
Journal Int J Med Robot
Publisher Wiley
Date 2021 Feb 3
PMID 33533112
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Background: A magnetic resonance image (MRI) guided robotic device for focussed ultrasound therapy of prostate cancer (PC) was developed. The device offers movement in 5 degrees of freedom (DOF) and uses a single-element transducer that operates at 3.2 MHz, has a diameter of 25 mm and focuses at 45 mm.

Methods: The MRI compatibility of the system was evaluated in a 1.5 T scanner. The ability of the transducer to create lesions was evaluated in laboratory and MRI settings, on ex vivo pork tissue and in vivo rabbit thigh tissue.

Results: Cavitational and thermal lesions were created on the excised pork tissue. In vivo experiments proved the efficacy of the system in ablating muscle tissue without damaging intervening areas.

Conclusions: The MRI compatible robotic system can be placed on the table of any commercial MRI scanner up to 7 T. The device has the ability of future use for transrectal focal therapy of PC with the patient in supine position.

Citing Articles

Magnetic Resonance Thermometry of Focused Ultrasound Using a Preclinical Focused Ultrasound Robotic System at 3T.

Filippou A, Evripidou N, Georgiou A, Georgiou L, Chrysanthou A, Ioannides C J Med Phys. 2025; 49(4):583-596.

PMID: 39926130 PMC: 11801101. DOI: 10.4103/jmp.jmp_133_24.


Magnetic Resonance Imaging Monitoring of Thermal Lesions Produced by Focused Ultrasound.

Antoniou A, Evripidou N, Nikolaou A, Georgiou A, Giannakou M, Chrysanthou A J Med Ultrasound. 2025; 32(4):297-308.

PMID: 39801543 PMC: 11717076. DOI: 10.4103/jmu.jmu_112_23.


Towards Design and Development of an MRI Conditional Robot to Enable Curvilinear Transperineal Prostate Biopsy.

Lafreniere S, Padasdao B, Konh B Int J Med Robot. 2024; 20(6):e70015.

PMID: 39641325 PMC: 11745543. DOI: 10.1002/rcs.70015.


A Comprehensive Review of Piezoelectric Ultrasonic Motors: Classifications, Characterization, Fabrication, Applications, and Future Challenges.

Naz S, Xu T Micromachines (Basel). 2024; 15(9).

PMID: 39337830 PMC: 11433840. DOI: 10.3390/mi15091170.


Effect of Magnetic Resonance Imaging on the Motion Accuracy of Magnetic Resonance Imaging-compatible Focused Ultrasound Robotic System.

Antoniou A, Evripidou N, Chrysanthou A, Georgiou L, Ioannides C, Spanoudes K J Med Phys. 2024; 49(2):203-212.

PMID: 39131431 PMC: 11309133. DOI: 10.4103/jmp.jmp_7_24.


References
1.
Napoli A, Anzidei M, De Nunzio C, Cartocci G, Panebianco V, De Dominicis C . Real-time magnetic resonance-guided high-intensity focused ultrasound focal therapy for localised prostate cancer: preliminary experience. Eur Urol. 2012; 63(2):395-8. DOI: 10.1016/j.eururo.2012.11.002. View

2.
Blana A, Murat F, Walter B, Thuroff S, Wieland W, Chaussy C . First analysis of the long-term results with transrectal HIFU in patients with localised prostate cancer. Eur Urol. 2007; 53(6):1194-201. DOI: 10.1016/j.eururo.2007.10.062. View

3.
Mearini L, Nunzi E, Giovannozzi S, Lepri L, Lolli C, Giannantoni A . Urodynamic evaluation after high-intensity focused ultrasound for patients with prostate cancer. Prostate Cancer. 2014; 2014:462153. PMC: 4052071. DOI: 10.1155/2014/462153. View

4.
Thuroff S, Chaussy C . Evolution and outcomes of 3 MHz high intensity focused ultrasound therapy for localized prostate cancer during 15 years. J Urol. 2013; 190(2):702-10. DOI: 10.1016/j.juro.2013.02.010. View

5.
Ficarra V, Zecchini Antoniolli S, Novara G, Parisi A, Fracalanza S, Martignoni G . Short-term outcome after high-intensity focused ultrasound in the treatment of patients with high-risk prostate cancer. BJU Int. 2006; 98(6):1193-8. DOI: 10.1111/j.1464-410X.2006.06561.x. View