» Articles » PMID: 36300728

Objective Burn Scar Assessment in Clinical Practice Using the Cutometer©: Introduction and Validation of a Standardized Measurement Protocol

Overview
Journal J Burn Care Res
Specialty Emergency Medicine
Date 2022 Oct 27
PMID 36300728
Authors
Affiliations
Soon will be listed here.
Abstract

An objective burn scar assessment is essential to informed therapeutic decision-making and to monitor scar development over time. However, widely employed scar rating scales show poor inter-rater reliability. For this study we developed a standardized measurement protocol for the Cutometer© applicable for objective burn scar assessment in everyday clinical practice. We developed a measurement protocol for the Cutometer© MPA 580 including a scar site relocation technique based on anatomical landmarks. The protocol emerged through several steps: Identifying key factors for valid and reliable measurements, preliminary testing, specification of technical details, refining the protocol and final testing. Consecutively, the protocol was validated for inter-rater reliability by assessing 34 burn scars in 17 patients by four clinicians and computing an Intra-class Correlation Coefficient (ICC). Parameter R0, representing scar pliability, was identified as the best suited output parameter yielding excellent inter-rater reliability for average measures (ICC 0.92 [95% CI 0.86; 0.96]) and acceptable reliability for single measures (ICC: 0.74 [0.61; 0.84]). The pressure applied on the measuring probe was identified as an influential confounding factor for reliable measurements. Rater gender did not influence reliability of measurements. The introduced standardized measurement protocol for the Cutometer© MPA 580 enables an objective and reliable burn scar assessment for clinical as well as research purposes.

Citing Articles

[Application of delayed replantation of degloving skin preserved at 4 in treatment of limb degloving injuries].

Xu Q, Xu J, Shen Y, Zhang C, Shen H, Huang T Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2025; 39(1):95-99.

PMID: 39848723 PMC: 11757951. DOI: 10.7507/1002-1892.202411057.


The transcription factor TCF4 regulates the miR-494-3p/THBS1 axis in the fibrosis of pathologic scars.

Lin G, Cao N, Wu J, Zheng M, Yang Z Arch Dermatol Res. 2025; 317(1):214.

PMID: 39786568 DOI: 10.1007/s00403-024-03692-9.


Efficacy and Complications Associated with Acellular Dermal Substitute Use in the Treatment of Acute Burns: A Systematic Review and Meta-Analysis.

Press I, Moiemen N, Ahmed Z Eur Burn J. 2024; 4(4):548-562.

PMID: 39600025 PMC: 11571822. DOI: 10.3390/ebj4040036.


Skin mechanical properties measured with skin elasticity measurement device in patients with lymphedema: Scoping review.

Fujimoto Y, Yuri Y, Tamiya H Skin Res Technol. 2024; 30(8):e13861.

PMID: 39096185 PMC: 11297454. DOI: 10.1111/srt.13861.


Intra- and inter-rater reliabilities of skin mechanical properties measured in healthy individuals using skin elasticity meter.

Fujimoto Y, Yuri Y, Kato Y, Kinoshita S, Tamiya H Ann Med. 2023; 55(2):2279747.

PMID: 37967233 PMC: 10653741. DOI: 10.1080/07853890.2023.2279747.

References
1.
Fearmonti R, Bond J, Erdmann D, Levinson H . A review of scar scales and scar measuring devices. Eplasty. 2010; 10:e43. PMC: 2890387. View

2.
Monteiro Rodrigues L, Fluhr J . EEMCO Guidance for the in vivo Assessment of Biomechanical Properties of the Human Skin and Its Annexes: Revisiting Instrumentation and Test Modes. Skin Pharmacol Physiol. 2019; 33(1):44-60. DOI: 10.1159/000504063. View

3.
Zuo K, Medina A, Tredget E . Important Developments in Burn Care. Plast Reconstr Surg. 2016; 139(1):120e-138e. DOI: 10.1097/PRS.0000000000002908. View

4.
Nedelec B, Correa J, Rachelska G, Armour A, LaSalle L . Quantitative measurement of hypertrophic scar: interrater reliability and concurrent validity. J Burn Care Res. 2008; 29(3):501-11. DOI: 10.1097/BCR.0b013e3181710881. View

5.
Shrout P, Fleiss J . Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979; 86(2):420-8. DOI: 10.1037//0033-2909.86.2.420. View