Microfluidic Microwave Biosensor Based on Biomimetic Materials for the Quantitative Detection of Glucose

Overview
Journal Sci Rep
Specialty Science
Date 2022 Sep 24
PMID 36153402
Authors
Affiliations
Soon will be listed here.
Abstract

This paper presents a microwave microfluidic biosensor for monitoring blood glucose levels. The glucose sensor is a triple ring microstrip patch antenna integrated with a biomimetic microfluidic device capable of measuring a fixed volume of glucose solution. The sensor was utilized to detect 50-500 mg/dL glucose solutions. The interaction of the glucose solution with the electromagnetic field on the patch's surface influences both the resonance frequency and the magnitude of reflection coefficient. The results indicate that the microfluidic device can reduce experimental error and enhance the correlation between glucose concentration, resonant frequency, and reflection coefficient. Finally, the microfluidic sensor had a sensitivity of 0.25 MHz/(mg/dL), a detection limit as low as 7.7 mg/dL, and correlation coefficients of resonance frequency and reflection coefficient with a glucose concentration of 0.996 and 0.984, respectively. The experiment on the sensor's stability verifies the sensor's excellent stability and rapid response (~ 150 ms). Consequently, the device can be used to differentiate the concentration of glucose solutions, as well as to detect blood glucose levels at an early stage.

Citing Articles

Clinical evaluation of a polarization-based optical noninvasive glucose sensing system.

Leung H, Gong C, Geiser L, Fivekiller E, Bui N, Vu T Sci Rep. 2025; 15(1):8877.

PMID: 40087308 DOI: 10.1038/s41598-025-92515-6.

References
1.
Yeon S, Seo M, Kim Y, Hong H, Chung T . Paper-based electrochromic glucose sensor with polyaniline on indium tin oxide nanoparticle layer as the optical readout. Biosens Bioelectron. 2022; 203:114002. DOI: 10.1016/j.bios.2022.114002. View

2.
Hu R, Stevenson A, Lowe C . An acoustic glucose sensor. Biosens Bioelectron. 2012; 35(1):425-428. DOI: 10.1016/j.bios.2012.02.026. View

3.
Ogurtsova K, Guariguata L, Barengo N, Ruiz P, Sacre J, Karuranga S . IDF diabetes Atlas: Global estimates of undiagnosed diabetes in adults for 2021. Diabetes Res Clin Pract. 2021; 183:109118. DOI: 10.1016/j.diabres.2021.109118. View

4.
Chen J, Zhou Q, Xu G, Wang R, Tai E, Xie L . Non-invasive blood glucose measurement of 95% certainty by pressure regulated Mid-IR. Talanta. 2019; 197:211-217. DOI: 10.1016/j.talanta.2019.01.034. View

5.
Govind G, Akhtar M . Design of an ELC resonator-based reusable RF microfluidic sensor for blood glucose estimation. Sci Rep. 2020; 10(1):18842. PMC: 7606440. DOI: 10.1038/s41598-020-75716-z. View