Clinically Compatible Subject-specific Dynamic Parallel Transmit Pulse Design for Homogeneous Fat Saturation and Water-excitation at 7T: Proof-of-concept for CEST MRI of the Brain

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2022 Sep 21
PMID 36128895
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the benefits and challenges of dynamic parallel transmit (pTx) pulses for fat saturation (FS) and water-excitation (WE), in the context of CEST MRI.

Methods: "Universal" k -points (for FS) and spiral non-selective (for WE) trajectories were optimized offline for flip angle (FA) homogeneity. Routines to optimize the pulse shape online, based on the subject's fields maps, were implemented (target FA of 110°/0° for FS, 0°/5° for WE at fat/water frequencies). The pulses were inserted in a CEST sequence with a pTx readout. The different fat suppression schemes and their effects on CEST contrasts were compared in 12 volunteers at 7T.

Results: With a 25%-shorter pulse duration, pTx FS largely improved the FA homogeneity (root-mean-square-error (RMSE) = 12.3° vs. 53.4° with circularly-polarized mode, at the fat frequency). However, the spectral selectivity was degraded mainly in the cerebellum and close to the sinuses (RMSE = 5.8° vs. 0.2° at the water frequency). Similarly, pTx WE showed a trade-off between FA homogeneity and spectral selectivity compared to pTx non-selective pulses (RMSE = 0.9° and 1.1° at the fat and water frequencies, vs. 4.6° and 0.5°). In the brain, CEST metrics were reduced by up to 31.9% at -3.3 ppm with pTx FS, suggesting a mitigated lipid-induced bias.

Conclusion: This clinically compatible implementation of dynamic pTx pulses improved the fat suppression homogeneity at 7T taking into account the subject-specific B heterogeneities online. This study highlights the lipid-induced biases on the CEST z-spectrum. The results are promising for body applications where B heterogeneities and fat are more substantial.

Citing Articles

Parallel-transmission spatial spectral pulse design with local specific absorption rate control: Demonstration for robust uniform water-selective excitation in the human brain at 7 T.

Shao X, Zhang Z, Ma X, Liu F, Guo H, Ugurbil K Magn Reson Med. 2024; 93(3):1238-1255.

PMID: 39481025 PMC: 11682926. DOI: 10.1002/mrm.30346.


Development and Validation of Four Different Methods to Improve MRI-CEST Tumor pH Mapping in Presence of Fat.

Gammaraccio F, Villano D, Irrera P, Anemone A, Carella A, Corrado A J Imaging. 2024; 10(7).

PMID: 39057737 PMC: 11277679. DOI: 10.3390/jimaging10070166.


Fast online spectral-spatial pulse design for subject-specific fat saturation in cervical spine and foot imaging at 1.5 T.

Eisen C, Liebig P, Herrler J, Ritter D, Levy S, Uder M MAGMA. 2024; 37(2):257-272.

PMID: 38366129 PMC: 10995033. DOI: 10.1007/s10334-024-01149-8.