At-home End-tidal Carbon Dioxide Measurement in Children with Invasive Home Mechanical Ventilation
Overview
Authors
Affiliations
Background: Carbon dioxide concentration trending is used in chronic management of children with invasive home mechanical ventilation (HMV) in clinical settings, but options for end-tidal carbon dioxide (EtCO ) monitoring at home are limited. We hypothesized that a palm-sized, portable endotracheal capnograph (PEC) that measures EtCO could be adapted for in-home use in children with HMV.
Methods: We evaluated the internal consistency of the PEC by calculating an intraclass correlation coefficient of three back-to-back breaths by children (0-17 years) at baseline health in the clinic. Pearson's correlation was calculated for PEC EtCO values with concurrent mean values of in-clinic EtCO and transcutaneous CO (TCM) capnometers. The Bland-Altman test determined their level of agreement. Qualitative interviews and surveys assessed usability and acceptability by family-caregivers at home.
Results: CO values were collected in awake children in varied activity levels and positions (N = 30). The intraclass correlation coefficient for the PEC was 0.95 (p < 0.05). The correlation between the PEC and in-clinic EtCO device was 0.85 with a mean difference of -3.8 mmHg and precision of ±1.1 mmHg. The correlation between the PEC and the clinic TCM device was 0.92 with a mean difference of 0.2 mmHg and precision of ±1.0. Family-caregivers (N = 10) trialed the PEC at home; all were able to obtain measurements at home while children were awake and sometimes asleep.
Conclusions: A portable, noninvasive device for measuring EtCO was feasible and acceptable, with values that trend similarly to currently in-practice, outpatient models. These devices may facilitate monitoring of EtCO at home in children with invasive HMV.
Ye C, Shablinski T, Shoaf S, Chung C, Bullock M Contemp Clin Trials Commun. 2025; 43():101396.
PMID: 39810842 PMC: 11731279. DOI: 10.1016/j.conctc.2024.101396.
Graham R, Amin R, Demirel N, Edel L, Lilien C, MacBean V Respir Res. 2024; 25(1):342.
PMID: 39285418 PMC: 11406763. DOI: 10.1186/s12931-024-02966-0.
An ultra-small integrated CO infrared gas sensor for wearable end-tidal CO monitoring.
Feng L, Liu Y, Wang Y, Zhou H, Wu M, Li T iScience. 2023; 26(11):108293.
PMID: 38026213 PMC: 10660086. DOI: 10.1016/j.isci.2023.108293.
Beidleman B, Figueiredo P, Landspurg S, Femling J, Williams J, Staab J J Appl Physiol (1985). 2023; 135(2):436-444.
PMID: 37318986 PMC: 10538982. DOI: 10.1152/japplphysiol.00216.2023.
At-home end-tidal carbon dioxide measurement in children with invasive home mechanical ventilation.
Foster C, Kwon S, Shah A, Hodgson C, Hird-McCorry L, Janus A Pediatr Pulmonol. 2022; 57(11):2735-2744.
PMID: 35959530 PMC: 9588689. DOI: 10.1002/ppul.26092.