Intrafractional Accuracy and Efficiency of a Surface Imaging System for Deep Inspiration Breath Hold During Ablative Gastrointestinal Cancer Treatment

Overview
Date 2022 Jul 30
PMID 35906884
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Beam gating with deep inspiration breath hold (DIBH) usually depends on some external surrogate to infer internal target movement, and the exact internal movement is unknown. In this study, we tracked internal targets and characterized residual motion during DIBH treatment, guided by a surface imaging system, for gastrointestinal cancer. We also report statistics on treatment time.

Methods And Materials: We included 14 gastrointestinal cancer patients treated with surface imaging-guided DIBH volumetrically modulated arc therapy, each with at least one radiopaque marker implanted near or within the target. They were treated in 25, 15, or 10 fractions. Thirteen patients received treatment for pancreatic cancer, and one underwent separate treatments for two liver metastases. The surface imaging system monitored a three-dimensional surface with ± 3 mm translation and ± 3° rotation threshold. During delivery, a kilovolt image was automatically taken every 20° or 40° gantry rotation, and the internal marker was identified from the image. The displacement and residual motion of the markers were calculated. To analyze the treatment efficiency, the treatment time of each fraction was obtained from the imaging and treatment timestamps in the record and verify system.

Results: Although the external surface was monitored and limited to ± 3 mm and ± 3°, significant residual internal target movement was observed in some patients. The range of residual motion was 3-21 mm. The average displacement for this cohort was 0-3 mm. In 19% of the analyzed images, the magnitude of the instantaneous displacement was > 5 mm. The mean treatment time was 17 min with a standard deviation of 4 min.

Conclusions: Precaution is needed when applying surface image guidance for gastrointestinal cancer treatment. Using it as a solo DIBH technique is discouraged when the correlation between internal anatomy and patient surface is limited. Real-time radiographic verification is critical for safe treatments.

Citing Articles

Clinical implementation and evaluation of stereotactic liver radiotherapy in inspiration breath-hold using nasal high-flow therapy and surface guidance.

Hazelaar C, Canters R, Kremer K, Lubken I, Vaassen F, Buijsen J Br J Radiol. 2024; 97(1164):1950-1958.

PMID: 39287019 PMC: 11573126. DOI: 10.1093/bjr/tqae177.


Optimizing Choice of Skin Surrogates for Surface-Guided Stereotactic Body Radiotherapy of Lung Lesions Using Four-Dimensional Computed Tomography.

Leipold V, Aleric I, Mlinaric M, Kosmina D, Stanic F, Kasabasic M Cancers (Basel). 2024; 16(13).

PMID: 39001420 PMC: 11240798. DOI: 10.3390/cancers16132358.


A feasibility trial of skin surface motion-gated stereotactic body radiotherapy for treatment of upper abdominal or lower thoracic targets using a novel O-ring gantry.

Kiser K, Schiff J, Laugeman E, Kim T, Green O, Hatscher C Clin Transl Radiat Oncol. 2023; 44:100692.

PMID: 38021090 PMC: 10652138. DOI: 10.1016/j.ctro.2023.100692.


Surface-Guided Radiotherapy: Can We Move on from the Era of Three-Point Markers to the New Era of Thousands of Points?.

Psarras M, Stasinou D, Stroubinis T, Protopapa M, Zygogianni A, Kouloulias V Bioengineering (Basel). 2023; 10(10).

PMID: 37892932 PMC: 10604452. DOI: 10.3390/bioengineering10101202.


Surface guidance compared with ultrasound-based monitoring and diaphragm position in cone-beam computed tomography during abdominal stereotactic radiotherapy in breath-hold.

Kaestner L, Streb L, Hetjens S, Buergy D, Sihono D, Fleckenstein J Phys Imaging Radiat Oncol. 2023; 27:100455.

PMID: 37720462 PMC: 10500027. DOI: 10.1016/j.phro.2023.100455.


References
1.
Bussels B, Goethals L, Feron M, Bielen D, Dymarkowski S, Suetens P . Respiration-induced movement of the upper abdominal organs: a pitfall for the three-dimensional conformal radiation treatment of pancreatic cancer. Radiother Oncol. 2003; 68(1):69-74. DOI: 10.1016/s0167-8140(03)00133-6. View

2.
Li G, Lu W, OGrady K, Yan I, Yorke E, Arriba L . A uniform and versatile surface-guided radiotherapy procedure and workflow for high-quality breast deep-inspiration breath-hold treatment in a multi-center institution. J Appl Clin Med Phys. 2022; 23(3):e13511. PMC: 8906224. DOI: 10.1002/acm2.13511. View

3.
Heerkens H, van Vulpen M, van den Berg C, Tijssen R, Crijns S, Molenaar I . MRI-based tumor motion characterization and gating schemes for radiation therapy of pancreatic cancer. Radiother Oncol. 2014; 111(2):252-7. DOI: 10.1016/j.radonc.2014.03.002. View

4.
Rong Y, Walston S, Welliver M, Chakravarti A, Quick A . Improving intra-fractional target position accuracy using a 3D surface surrogate for left breast irradiation using the respiratory-gated deep-inspiration breath-hold technique. PLoS One. 2014; 9(5):e97933. PMC: 4031138. DOI: 10.1371/journal.pone.0097933. View

5.
Remouchamps V, Letts N, Vicini F, Sharpe M, Kestin L, Chen P . Initial clinical experience with moderate deep-inspiration breath hold using an active breathing control device in the treatment of patients with left-sided breast cancer using external beam radiation therapy. Int J Radiat Oncol Biol Phys. 2003; 56(3):704-15. DOI: 10.1016/s0360-3016(03)00010-5. View