Phantom-based Correction for Standardization of Myocardial Native T1 and Extracellular Volume Fraction in Healthy Subjects at 3-Tesla Cardiac Magnetic Resonance Imaging
Overview
Affiliations
Objectives: To investigate the effect of the phantom-based correction method for standardizing myocardial native T1 and extracellular volume fraction (ECV) in healthy subjects.
Methods: Seventy-one healthy asymptomatic adult (≥ 20 years) volunteers of five different age groups (34 men and 37 women, 45.5 ± 15.5 years) were prospectively enrolled in three academic hospitals. Cardiac MRI including Modified Look - Locker Inversion recovery T1 mapping sequence was performed using a 3-Tesla system with a different type of scanner for each hospital. Native T1 and ECV were measured in the short-axis T1 map and analyzed for mean values of the 16 entire segments. The myocardial T1 value of each subject was corrected based on the site-specific equation derived from the T1 Mapping and ECV Standardization phantom. The global native T1 and ECV were compared between institutions before and after phantom-based correction, and the variation in native T1 and ECV among institutions was assessed using a coefficient of variation (CoV).
Results: The global native T1 value significantly differed between the institutions (1198.7 ± 32.1 ms, institution A; 1217.7 ± 39.9 ms, institution B; 1232.7 ± 31.1 ms, institution C; p = 0.002), but the mean ECV did not (26.6-27.5%, p = 0.355). After phantom-based correction, the global native T1 and ECV were 1289.7 ± 32.4 ms and 25.0 ± 2.7%, respectively, and CoV for native T1 between the three institutions decreased from 3.0 to 2.5%. The corrected native T1 value did not significantly differ between institutions (1284.5 ± 31.5 ms, institution A; 1296.5 ± 39.1 ms, institution B; 1291.3 ± 29.3 ms, institution C; p = 0.440), and neither did the ECV (24.4-25.9%, p = 0.078).
Conclusions: The phantom-based correction method can provide standardized reference T1 values in healthy subjects.
Key Points: • After phantom-based correction, the global native T1 of 16 entire myocardial segments on 3-T cardiac MRI is 1289.4 ± 32.4 ms, and the extracellular volume fraction was 25.0 ± 2.7% for healthy subjects. • After phantom - based correction was applied, the differences in the global native T1 among institutions became insignificant, and the CoV also decreased from 3.0 to 2.5%.
Galbusera F, Cina A, ORiordan D, Vitale J, Loibl M, Fekete T Eur Spine J. 2024; 33(11):4092-4103.
PMID: 39212711 DOI: 10.1007/s00586-024-08463-8.
Kim H, Yang Y, Han K, Kim P, Choi B, Kim J Quant Imaging Med Surg. 2023; 13(10):6750-6760.
PMID: 37869306 PMC: 10585511. DOI: 10.21037/qims-23-375.
Chang S, Han K, Lee S, Yang Y, Kim P, Choi B Korean J Radiol. 2022; 23(12):1251-1259.
PMID: 36447413 PMC: 9747268. DOI: 10.3348/kjr.2022.0496.
Myocardial T2* Imaging at 3T and 1.5T: A Pilot Study with Phantom and Normal Myocardium.
Chang S, Park J, Yang Y, Beck K, Kim P, Choi B J Cardiovasc Dev Dis. 2022; 9(8).
PMID: 36005435 PMC: 9410052. DOI: 10.3390/jcdd9080271.