Objective: The purpose of this study was to investigate the value of the lateral point of articular surface of distal tibia (LADT) for anatomical alignment in total knee arthroplasty.

Methods: We reconstructed 148 three-dimensional pre-arthritic tibias and measured the tibial component inclination angle corresponding to the distal landmark of LADT. A retrospective study included 81 TKA recipients divided into the AA group and MA group. Clinical assessments including ROM, HSS, WOMAC, satisfaction for surgery, and radiological assessment were evaluated at one-year follow-up.

Results: The tibial component varus angle corresponding to the distal landmark of LADT in the male and female groups were 3.4 ± 0.3° (2.6~4.2°) and 3.2 ± 0.3° (2.3~4.0°), respectively ( <0.05). Using LADT as the distal landmark for extramedullary tibial cutting guidance, the medial proximal tibia angle (MPTA) of the AA group was 87.0±1.2° (85.0~90.0°), and the AA and MA technique showed no difference in improvement in postoperative knee functional recovery at final follow-up.

Conclusions: This study preliminarily indicated that LADT can be a reliable and economical landmark for coronal plane alignment of the tibial component.

Citing Articles

The distal tibiofibular syndesmosis is a reliable landmark for 3° varus tibial resection in total knee arthroplasty: a radiological evaluation on 1296 cases.

Perelli S, Costa G, Russo A, Hinarejos P, Torres-Claramunt R, Sanchez-Soler J Arch Orthop Trauma Surg. 2023; 144(2):879-885.

PMID: 37864591 DOI: 10.1007/s00402-023-05099-z.

References
1.
Nam D, Nunley R, Barrack R . Patient dissatisfaction following total knee replacement: a growing concern?. Bone Joint J. 2014; 96-B(11 Supple A):96-100. DOI: 10.1302/0301-620X.96B11.34152. View

2.
Thienpont E, Paternostre F, Van Wymeersch C . The indirect cost of Patient-Specific Instruments. Acta Orthop Belg. 2015; 81(3):462-70. View

3.
Nizam I, Batra A . Accuracy of bone resection in total knee arthroplasty using CT assisted-3D printed patient specific cutting guides. SICOT J. 2018; 4:29. PMC: 6047362. DOI: 10.1051/sicotj/2018032. View

4.
Berend M, Ritter M, Meding J, Faris P, Keating E, Redelman R . Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res. 2004; (428):26-34. DOI: 10.1097/01.blo.0000148578.22729.0e. View

5.
Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C . A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Joint J. 2013; 95-B(3):354-9. DOI: 10.1302/0301-620X.95B3.29903. View