» Articles » PMID: 35361435

Microfluidic Aptasensor POC Device for Determination of Whole Blood Potassium

Overview
Journal Anal Chim Acta
Publisher Elsevier
Specialty Chemistry
Date 2022 Apr 1
PMID 35361435
Authors
Affiliations
Soon will be listed here.
Abstract

An integrated microfluidic Au nanoparticle (AuNP) aptasensor device is proposed for monitoring the concentration of potassium (K) ions in the bloodstream of patients with chronic kidney disease (CKD). In the proposed detection device, the AuNPs in the AuNP/aptamer complex are displaced by the serum K ions and react with NaCl to produce a color change in the detection region from which the K ion concentration is then inversely derived. The microfluidic device comprises two main components, namely an AuNP aptasensor PMMA (Poly(methyl methacrylate))/paper-microchip and a colorimetric analysis system for the quantitative detection of K ion concentration in whole blood. The functions of PMMA/paper microchips include reagent storage, K ion/aptamer reaction, and separation of serum from whole blood samples (blood filter). Experimental results show that the microfluidic device provides a linear response over the K ion concentration in range of 0.05-9 mM in artificial serum and has a detection limit (LOD) of 0.01 mM. Moreover, the detection results obtained for the 137 whole blood and 287 serum samples of CKD patients are very consistent (R = 0.968 and R = 0.980) with the measurement results obtained using an ion-selective electrodes (ISE) method. Results confirm that the current microfluidic aptasensor device provides a highly-sensitive and convenient method for performing the point-of-care (POC) monitoring of the whole blood K ion concentration.

Citing Articles

Advancements in Biosensors for Lipid Peroxidation and Antioxidant Protection in Food: A Critical Review.

Daci M, Berisha L, Mercatante D, Rodriguez-Estrada M, Jin Z, Huang Y Antioxidants (Basel). 2025; 13(12.

PMID: 39765813 PMC: 11672933. DOI: 10.3390/antiox13121484.


Microfluidic Detection Platform for Determination of Ractopamine in Food.

Yu C, Huang K, Chen T, Liu C, Fu L Biosensors (Basel). 2024; 14(10).

PMID: 39451675 PMC: 11506807. DOI: 10.3390/bios14100462.


Rapid Microfluidic Immuno-Biosensor Detection System for the Point-of-Care Determination of High-Sensitivity Urinary C-Reactive Protein.

Chen S, Lu S, Tseng C, Huang K, Chen T, Fu L Biosensors (Basel). 2024; 14(6).

PMID: 38920587 PMC: 11201708. DOI: 10.3390/bios14060283.


Microfluidic Blood Separation: Key Technologies and Critical Figures of Merit.

Torres-Castro K, Acuna-Umana K, Lesser-Rojas L, Reyes D Micromachines (Basel). 2023; 14(11).

PMID: 38004974 PMC: 10672873. DOI: 10.3390/mi14112117.


Aptamer-Based Point-of-Care Devices: Emerging Technologies and Integration of Computational Methods.

Aslan Y, Atabay M, Chowdhury H, Gokturk I, Saylan Y, Inci F Biosensors (Basel). 2023; 13(5).

PMID: 37232930 PMC: 10216294. DOI: 10.3390/bios13050569.