» Articles » PMID: 35323412

An Electrochemical Impedance Spectroscopy-Based Aptasensor for the Determination of SARS-CoV-2-RBD Using a Carbon Nanofiber-Gold Nanocomposite Modified Screen-Printed Electrode

Overview
Specialty Biotechnology
Date 2022 Mar 24
PMID 35323412
Authors
Affiliations
Soon will be listed here.
Abstract

Worldwide, human health is affected by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Hence, the fabrication of the biosensors to diagnose SARS-CoV-2 is critical. In this paper, we report an electrochemical impedance spectroscopy (EIS)-based aptasensor for the determination of the SARS-CoV-2 receptor-binding domain (SARS-CoV-2-RBD). For this purpose, the carbon nanofibers (CNFs) were first decorated with gold nanoparticles (AuNPs). Then, the surface of the carbon-based screen-printed electrode (CSPE) was modified with the CNF-AuNP nanocomposite (CSPE/CNF-AuNP). After that, the thiol-terminal aptamer probe was immobilized on the surface of the CSPE/CNF-AuNP. The surface coverage of the aptamer was calculated to be 52.8 pmol·cm. The CSPE/CNF-AuNP/Aptamer was then used for the measurement of SARS-CoV-2-RBD by using the EIS method. The obtained results indicate that the signal had a linear-logarithmic relationship in the range of 0.01-64 nM with a limit of detection of 7.0 pM. The proposed aptasensor had a good selectivity to SARS-CoV-2-RBD in the presence of human serum albumin; human immunoglobulins G, A, and M, hemagglutinin, and neuraminidase. The analytical performance of the aptasensor was studied in human saliva samples. The present study indicates a practical application of the CSPE/CNF-AuNP/Aptamer for the determination of SARS-CoV-2-RBD in human saliva samples with high sensitivity and accuracy.

Citing Articles

Advances in textile-based microfluidics for biomolecule sensing.

Milic L, Zambry N, Ibrahim F, Petrovic B, Kojic S, Thiha A Biomicrofluidics. 2024; 18(5):051502.

PMID: 39296324 PMC: 11410389. DOI: 10.1063/5.0222244.


Sensing Characteristics of SARS-CoV-2 Spike Protein Using Aptamer-Functionalized Si-Based Electrolyte-Gated Field-Effect Transistor (EGT).

Shin S, Kim S, Choi W, Do J, Son J, Kim K Biosensors (Basel). 2024; 14(3).

PMID: 38534231 PMC: 10967824. DOI: 10.3390/bios14030124.


Advancement in Paper-Based Electrochemical Biosensing and Emerging Diagnostic Methods.

Benjamin S, de Lima F, do Nascimento V, de Andrade G, Barreto Oria R Biosensors (Basel). 2023; 13(7).

PMID: 37504088 PMC: 10377443. DOI: 10.3390/bios13070689.


Electrochemical investigation of hydroxyapatite-lanthanum strontium cobalt ferrite composites (HA-LSCF) for SARS-CoV-2 aptasensors.

Hartati Y, Devi M, Irkham , Zulqaidah S, Noviyanti A, Rochani S RSC Adv. 2023; 13(29):20209-20216.

PMID: 37416913 PMC: 10321058. DOI: 10.1039/d3ra01531a.


Imunocapture Magnetic Beads Enhanced and Ultrasensitive CRISPR-Cas13a-Assisted Electrochemical Biosensor for Rapid Detection of SARS-CoV-2.

Han Y, Li F, Yang L, Guo X, Dong X, Niu M Biosensors (Basel). 2023; 13(6).

PMID: 37366962 PMC: 10296353. DOI: 10.3390/bios13060597.


References
1.
Fabiani L, Saroglia M, Galata G, De Santis R, Fillo S, Luca V . Magnetic beads combined with carbon black-based screen-printed electrodes for COVID-19: A reliable and miniaturized electrochemical immunosensor for SARS-CoV-2 detection in saliva. Biosens Bioelectron. 2020; 171:112686. PMC: 7833515. DOI: 10.1016/j.bios.2020.112686. View

2.
Yakoh A, Pimpitak U, Rengpipat S, Hirankarn N, Chailapakul O, Chaiyo S . Paper-based electrochemical biosensor for diagnosing COVID-19: Detection of SARS-CoV-2 antibodies and antigen. Biosens Bioelectron. 2020; 176:112912. PMC: 7746088. DOI: 10.1016/j.bios.2020.112912. View

3.
Kumeria T, Santos A, Losic D . Ultrasensitive nanoporous interferometric sensor for label-free detection of gold(III) ions. ACS Appl Mater Interfaces. 2013; 5(22):11783-90. DOI: 10.1021/am403465x. View

4.
Nowak E, Wisla-Swider A, Khachatryan G, Fiedorowicz M, Danel K . Possible sensor applications of selected DNA-surfactant complexes. Eur Biophys J. 2019; 48(4):371-381. DOI: 10.1007/s00249-019-01367-2. View

5.
Mojsoska B, Larsen S, Olsen D, Madsen J, Brandslund I, Alatraktchi F . Rapid SARS-CoV-2 Detection Using Electrochemical Immunosensor. Sensors (Basel). 2021; 21(2). PMC: 7827295. DOI: 10.3390/s21020390. View