Individual Pulse Monitoring and Dose Control System for Pre-clinical Implementation of FLASH-RT

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2022 Mar 21
PMID 35313290
Authors
Affiliations
Soon will be listed here.
Abstract

Existing ultra-high dose rate (UHDR) electron sources lack dose rate independent dosimeters and a calibrated dose control system for accurate delivery. In this study, we aim to develop a custom single-pulse dose monitoring and a real-time dose-based control system for a FLASH enabled clinical linear accelerator (Linac).A commercially available point scintillator detector was coupled to a gated integrating amplifier and a real-time controller for dose monitoring and feedback control loop. The controller was programmed to integrate dose for each radiation pulse and stop the radiation beam when the prescribed dose was delivered. Additionally, the scintillator was mounted in a solid water phantom and placed underneath mice skin fordose monitoring. The scintillator was characterized in terms of its radiation stability, mean dose-rate (Ḋm), and dose per pulse () dependence.Theexhibited a consistent ramp-up period across ∼4-5 pulse. The plastic scintillator was shown to be linear withḊm(40-380 Gy s) and(0.3-1.3 Gy Pulse) to within +/- 3%. However, the plastic scintillator was subject to significant radiation damage (16%/kGy) for the initial 1 kGy and would need to be calibrated frequently. Pulse-counting control was accurately implemented with one-to-one correspondence between the intended and the actual delivered pulses. The dose-based control was sufficient to gate on any pulse of the Linac.dosimetry monitoring with a 1 cm circular cut-out revealed that during the ramp-up period, the averagewas ∼0.045 ± 0.004 Gy Pulse, whereas after the ramp-up it stabilized at 0.65 ± 0.01 Gy Pulse.The tools presented in this study can be used to determine the beam parameter space pertinent to the FLASH effect. Additionally, this study is the first instance of real-time dose-based control for a modified Linac at ultra-high dose rates, which provides insight into the tool required for future clinical translation of FLASH-RT.

Citing Articles

A comprehensive investigation of the performance of a commercial scintillator system for applications in electron FLASH radiotherapy.

Liu K, Holmes S, Schuler E, Beddar S Med Phys. 2024; 51(6):4504-4512.

PMID: 38507253 PMC: 11147715. DOI: 10.1002/mp.17030.


Characterization of a novel time-resolved, real-time scintillation dosimetry system for ultra-high dose rate radiation therapy applications.

Baikalov A, Tho D, Liu K, Bartzsch S, Beddar S, Schuler E ArXiv. 2024; .

PMID: 38495573 PMC: 10942482.


Beam control system and output fine-tuning for safe and precise delivery of FLASH radiotherapy at a clinical linear accelerator.

Konradsson E, Wahlqvist P, Thoft A, Blad B, Back S, Ceberg C Front Oncol. 2024; 14:1342488.

PMID: 38304871 PMC: 10830783. DOI: 10.3389/fonc.2024.1342488.


Technical note: High-dose and ultra-high dose rate (UHDR) evaluation of Al O :C optically stimulated luminescent dosimeter nanoDots and powdered LiF:Mg,Ti thermoluminescent dosimeters for radiation therapy applications.

Liu K, Velasquez B, Schuler E Med Phys. 2023; 51(3):2311-2319.

PMID: 37991111 PMC: 10939935. DOI: 10.1002/mp.16832.


Characterization of a diode dosimeter for UHDR FLASH radiotherapy.

Rahman M, Kozelka J, Hildreth J, Schonfeld A, Sloop A, Ashraf M Med Phys. 2023; 50(9):5875-5883.

PMID: 37249058 PMC: 11748214. DOI: 10.1002/mp.16474.


References
1.
Jaccard M, Petersson K, Buchillier T, Germond J, Duran M, Vozenin M . High dose-per-pulse electron beam dosimetry: Usability and dose-rate independence of EBT3 Gafchromic films. Med Phys. 2016; 44(2):725-735. DOI: 10.1002/mp.12066. View

2.
Pettinato S, Girolami M, Olivieri R, Stravato A, Caruso C, Salvatori S . A Diamond-Based Dose-per-Pulse X-ray Detector for Radiation Therapy. Materials (Basel). 2021; 14(18). PMC: 8466252. DOI: 10.3390/ma14185203. View

3.
Jursinic P . Dependence of diode sensitivity on the pulse rate of delivered radiation. Med Phys. 2013; 40(2):021720. DOI: 10.1118/1.4788763. View

4.
Jaccard M, Duran M, Petersson K, Germond J, Liger P, Vozenin M . High dose-per-pulse electron beam dosimetry: Commissioning of the Oriatron eRT6 prototype linear accelerator for preclinical use. Med Phys. 2017; 45(2):863-874. DOI: 10.1002/mp.12713. View

5.
Frelin A, Fontbonne J, Ban G, Colin J, Labalme M, Batalla A . Spectral discrimination of Cerenkov radiation in scintillating dosimeters. Med Phys. 2005; 32(9):3000-6. DOI: 10.1118/1.2008487. View