Technical Note: Time-gating to Medical Linear Accelerator Pulses: Stray Radiation Detector

Overview
Journal Med Phys
Specialty Biophysics
Date 2018 Nov 30
PMID 30488442
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: CCD cameras are employed to image scintillation and Cherenkov radiation in external beam radiotherapy. This is achieved by gating the camera to the linear accelerator (Linac) output. A direct output signal line from the linac is not always accessible and even in cases where such a signal is accessible, a physical wire connected to the output port can potentially alter Linac performance through electrical feedback. A scintillating detector for stray radiation inside the Linac room was developed to remotely time-gate to linac pulses for camera-based dosimetry.

Methods: A scintillator coupled silicon photomultiplier detector was optimized and systematically tested for location sensitivity and for use with both x rays and electron beams, at different energies and field sizes. Cherenkov radiation emitted due to static photon beams was captured using the remote trigger and compared to the images captured using a wired trigger. The issue of false-positive event detection, due to additional neutron activated products with high energy beams, was addressed.

Results: The designed circuit provided voltage >2.5 V even for distances up to 3 m from the isocenter with a 6 MV, 5 × 5 cm beam, using a Ø3 × 20 mm Bi Ge O (BGO) crystal. With a larger scintillator size, the detector could be placed even beyond 3 m distance. False-positive triggering was reduced by a coincidence detection scheme. Negligible fluctuations were observed in time-gated imaging of Cherenkov intensity emitted from a water phantom, when comparing directly connected vs this remote triggering approach.

Conclusion: The remote detector provides untethered synchronization to linac pulses. It is especially useful for remote Cherenkov imaging or remote scintillator dosimetry imaging during radiotherapeutic procedures when a direct line signal is not accessible.

Citing Articles

Technical note: Visual, rapid, scintillation point dosimetry for in vivo MV photon beam radiotherapy treatments.

Decker S, Bruza P, Zhang R, Williams B, Jarvis L, Pogue B Med Phys. 2024; 51(8):5754-5763.

PMID: 38598093 PMC: 11321933. DOI: 10.1002/mp.17071.


Characterization of Cherenkov imaging parameters and positional constraints on an O-ring linear accelerator.

Alexander D, Majji S, Jermyn M, Byrd B, Bruza P, Li T Phys Med Biol. 2023; 68(20).

PMID: 37757840 PMC: 10693929. DOI: 10.1088/1361-6560/acfdf2.


Color-resolved Cherenkov imaging allows for differential signal detection in blood and melanin content.

Wickramasinghe V, Decker S, Streeter S, Sloop A, Petusseau A, Alexander D J Biomed Opt. 2023; 28(3):036005.

PMID: 36923987 PMC: 10008915. DOI: 10.1117/1.JBO.28.3.036005.


Performance comparison of quantitative metrics for analysis of in vivo Cherenkov imaging incident detection during radiotherapy.

Decker S, Alexander D, Bruza P, Zhang R, Chen E, Jarvis L Br J Radiol. 2022; 95(1137):20211346.

PMID: 35834415 PMC: 10996952. DOI: 10.1259/bjr.20211346.


One Year of Clinic-Wide Cherenkov Imaging for Discovery of Quality Improvement Opportunities in Radiation Therapy.

Alexander D, Decker S, Jermyn M, Bruza P, Zhang R, Chen E Pract Radiat Oncol. 2022; 13(1):71-81.

PMID: 35777728 PMC: 10984217. DOI: 10.1016/j.prro.2022.06.009.


References
1.
Fischer H, Tabot B, Poppe B . Activation processes in a medical linear accelerator and spatial distribution of activation products. Phys Med Biol. 2006; 51(24):N461-6. DOI: 10.1088/0031-9155/51/24/N02. View

2.
Howell R, Kry S, Burgett E, Hertel N, Followill D . Secondary neutron spectra from modern Varian, Siemens, and Elekta linacs with multileaf collimators. Med Phys. 2009; 36(9):4027-38. PMC: 2738742. DOI: 10.1118/1.3159300. View

3.
Glaser A, Zhang R, Davis S, Gladstone D, Pogue B . Time-gated Cherenkov emission spectroscopy from linear accelerator irradiation of tissue phantoms. Opt Lett. 2012; 37(7):1193-5. PMC: 3698962. DOI: 10.1364/OL.37.001193. View

4.
Jarvis L, Zhang R, Gladstone D, Jiang S, Hitchcock W, Friedman O . Cherenkov video imaging allows for the first visualization of radiation therapy in real time. Int J Radiat Oncol Biol Phys. 2014; 89(3):615-22. DOI: 10.1016/j.ijrobp.2014.01.046. View