» Articles » PMID: 35083979

Adverse Events Associated with AXIOS Stents: Insights from the Manufacturer and User Facility Device Experience Database

Overview
Date 2022 Jan 27
PMID 35083979
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Objectives: The AXIOS stent is indicated for transgastric or transduodenal drainage of symptomatic pancreatic pseudocysts and symptomatic walled off necrosis. The AXIOS stent functions as a conduit which allows solid and liquid pancreatic fluid collections (PFC) contents to pass into the luminal GI tract and also allows the passage of standard and therapeutic endoscopes into the PFC to perform endoscopic debridement. We aim to investigate the number and type of complications associated with AXIOS stents.

Materials And Methods: We analyzed postmarketing surveillance data from the Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database January 2016 to February 2021.

Results: During the study period, approximately 588 reports with 579 device issues and 250 patient complications were identified. Most device complications were due to stent positioning problems or stent malpositioning (n = 206; 35.6%), followed by stent migration (n = 72; 12.4%), premature deployment (n = 61; 10.5%), material integrity (n = 56; 9.6%), deployment failure (n = 47; 8.1%), and difficulty removing the stent (n = 45; 7.7%). The most reported patient adverse events were hemorrhage/bleeding (n = 81; 32.4%), perforation (n = 26; 10.4%), pain (n = 22; 8.8%), unspecified infection (n = 20; 8.0%), and death (n = 17; 6.8%).

Conclusions: Findings from the MAUDE database highlight patient and device complications which endoscopists should be aware of before AXIOS stent placement.

Citing Articles

A novel lumen-apposing metal stent for endoscopic drainage of symptomatic pancreatic fluid collections: a retrospective study.

Xu N, Li L, Su S, Zhao D, Xiang J, Wang P Endosc Ultrasound. 2024; 13(1):40-45.

PMID: 38947119 PMC: 11213609. DOI: 10.1097/eus.0000000000000039.


Small bowel obstruction due to a migrated pyloric stent.

Agolia J, Wang S, Fisher A, Bryan J, Knowlton L Trauma Surg Acute Care Open. 2024; 9(1):e001443.

PMID: 38756695 PMC: 11097799. DOI: 10.1136/tsaco-2024-001443.


Assessing the Reproducibility of Research Based on the Food and Drug Administration Manufacturer and User Facility Device Experience Data.

Li X, Feng Y, Gong Y, Chen Y J Patient Saf. 2024; 20(5):e45-e58.

PMID: 38470959 PMC: 11636620. DOI: 10.1097/PTS.0000000000001220.


EUS-guided interventional therapies for pancreatic diseases.

Xu R, Zhang K, Ge N, Sun S Front Med (Lausanne). 2024; 10:1329676.

PMID: 38259846 PMC: 10801084. DOI: 10.3389/fmed.2023.1329676.


Clinical application and follow-up of electrocautery-enhanced lumen-apposing metal stents for endoscopic ultrasonography-guided interventions.

She Y, Ge N Ann Med. 2023; 55(2):2282748.

PMID: 37967551 PMC: 10653748. DOI: 10.1080/07853890.2023.2282748.


References
1.
Walter D, Will U, Sanchez-Yague A, Brenke D, Hampe J, Wollny H . A novel lumen-apposing metal stent for endoscopic ultrasound-guided drainage of pancreatic fluid collections: a prospective cohort study. Endoscopy. 2014; 47(1):63-7. DOI: 10.1055/s-0034-1378113. View

2.
Singh Dhindsa B, Mashiana H, Dhaliwal A, Mohan B, Jayaraj M, Sayles H . EUS-guided biliary drainage: A systematic review and meta-analysis. Endosc Ultrasound. 2020; 9(2):101-109. PMC: 7279084. DOI: 10.4103/eus.eus_80_19. View

3.
Chandan S, Khan S, Mohan B, Shah A, Bilal M, Ramai D . EUS-guided gastroenterostomy versus enteral stenting for gastric outlet obstruction: Systematic review and meta-analysis. Endosc Int Open. 2021; 9(3):E496-E504. PMC: 7899789. DOI: 10.1055/a-1341-0788. View

4.
Fugazza A, Sethi A, Trindade A, Troncone E, Devlin J, Khashab M . International multicenter comprehensive analysis of adverse events associated with lumen-apposing metal stent placement for pancreatic fluid collection drainage. Gastrointest Endosc. 2019; 91(3):574-583. DOI: 10.1016/j.gie.2019.11.021. View

5.
Siddiqui A, Kowalski T, Loren D, Khalid A, Soomro A, Mazhar S . Fully covered self-expanding metal stents versus lumen-apposing fully covered self-expanding metal stent versus plastic stents for endoscopic drainage of pancreatic walled-off necrosis: clinical outcomes and success. Gastrointest Endosc. 2016; 85(4):758-765. DOI: 10.1016/j.gie.2016.08.014. View