How to Validate Deployment of Coronary Stents: Strategies and Limitations in the Choice of Comparator

Overview
Date 2022 Jan 20
PMID 35047942
Authors
Affiliations
Soon will be listed here.
Abstract

This study aims at proposing and discussing useful indications to all those who need to validate a numerical model of coronary stent deployment. The proof of the reliability of a numerical model is becoming of paramount importance in the era of trials. Recently, the ASME V&V Standard Committee for medical devices prepared the V&V 40 standard document that provides a framework that guides users in establishing and assessing the relevance and adequacy of verification and validation activities performed for proving the credibility of models. To the knowledge of the authors, only a few examples of the application of the V&V 40 framework to medical devices are available in the literature, but none about stents. Specifically, in this study, the authors wish to emphasize the choice of a relevant set of experimental activities to provide data for the validation of computational models aiming to predict coronary stent deployment. Attention is focused on the use of 3D-printed mock vessels in the validation plan, which could allow evaluating aspects of clinical relevance in a representative but controlled environment.

Citing Articles

Investigating Balloon-Vessel Contact Pressure Patterns in Angioplasty: In Silico Insights for Drug-Coated Balloons.

Stratakos E, Antonini L, Poletti G, Berti F, Tzafriri A, Petrini L Ann Biomed Eng. 2023; 51(12):2908-2922.

PMID: 37751027 PMC: 10632265. DOI: 10.1007/s10439-023-03359-y.


Recommendations for finite element modelling of nickel-titanium stents-Verification and validation activities.

Bernini M, Hellmuth R, Dunlop C, Ronan W, Vaughan T PLoS One. 2023; 18(8):e0283492.

PMID: 37556457 PMC: 10411813. DOI: 10.1371/journal.pone.0283492.


Relevant Choices Affecting the Fatigue Analysis of Ni-Ti Endovascular Devices.

Berti F, Brambilla A, Pennati G, Petrini L Materials (Basel). 2023; 16(8).

PMID: 37110014 PMC: 10146368. DOI: 10.3390/ma16083178.


Fabrication of Compliant and Transparent Hollow Cerebral Vascular Phantoms for In Vitro Studies Using 3D Printing and Spin-Dip Coating.

Bisighini B, Di Giovanni P, Scerrati A, Trovalusci F, Vesco S Materials (Basel). 2023; 16(1).

PMID: 36614505 PMC: 9821401. DOI: 10.3390/ma16010166.


Application of Platform for the Development and Optimization of Fully Bioresorbable Vascular Scaffold Designs.

Milosevic M, Anic M, Nikolic D, Geroski V, Milicevic B, Kojic M Front Med Technol. 2022; 3:724062.

PMID: 35047953 PMC: 8757700. DOI: 10.3389/fmedt.2021.724062.

References
1.
Blair R, Dunne N, Lennon A, Menary G . Characterisation and constitutive modelling of biaxially stretched poly(L-lactic acid) sheet for application in coronary stents. J Mech Behav Biomed Mater. 2019; 97:346-354. DOI: 10.1016/j.jmbbm.2019.05.039. View

2.
Pathmanathan P, Gray R . Validation and Trustworthiness of Multiscale Models of Cardiac Electrophysiology. Front Physiol. 2018; 9:106. PMC: 5818422. DOI: 10.3389/fphys.2018.00106. View

3.
Medrano-Gracia P, Ormiston J, Webster M, Beier S, Ellis C, Wang C . A Study of Coronary Bifurcation Shape in a Normal Population. J Cardiovasc Transl Res. 2016; 10(1):82-90. PMC: 5323506. DOI: 10.1007/s12265-016-9720-2. View

4.
De Bock S, Iannaccone F, De Santis G, De Beule M, Van Loo D, Devos D . Virtual evaluation of stent graft deployment: a validated modeling and simulation study. J Mech Behav Biomed Mater. 2012; 13:129-39. DOI: 10.1016/j.jmbbm.2012.04.021. View

5.
Larrabide I, Kim M, Augsburger L, Villa-Uriol M, Rufenacht D, Frangi A . Fast virtual deployment of self-expandable stents: method and in vitro evaluation for intracranial aneurysmal stenting. Med Image Anal. 2010; 16(3):721-30. DOI: 10.1016/j.media.2010.04.009. View