Materials And Methods: Forty sterile endodontic rotary files were selected and distributed into the following study groups: A: 25.06 double S-shaped cross-section NiTi alloy endodontic rotary files (Mtwo) ( = 10); B: 20.04 rectangular cross-section NiTi alloy endodontic rotary files (T Pro E1) ( = 10); C: 25.04 convex triangular cross-section NiTi alloy endodontic rotary files (T Pro E2) ( = 10); and D: 25.06 triangular cross-section NiTi alloy endodontic rotary files (T Pro E4) ( = 10). A cyclic fatigue device was used to conduct the static cyclic fatigue tests with stainless steel artificial root canal systems with 200 µm and 250 µm apical diameter, 60° curvature angle, 3 mm radius of curvature, 20 mm length, and 4% and 8% taper. The results were analyzed using the ANOVA test and Weibull statistical analysis.

Results: All the pairwise comparisons presented statistically significant differences between the time to failure and number of cycles to failure for the cross-section design study groups ( < 0.001).

Conclusions: the double S-shaped cross-section of Mtwo NiTi endodontic files shows higher cyclic fatigue resistance than the rectangular cross-section of T Pro E1 NiTi endodontic files, the convex triangular cross-section of T Pro E2 NiTi endodontic files, and the triangular cross-section of T Pro E4 NiTi endodontic files.

Citing Articles

Cyclic Fatigue of Different Reciprocating Endodontic Instruments Using Matching Artificial Root Canals at Body Temperature In Vitro.

Burklein S, Massmann P, Schafer E, Donnermeyer D Materials (Basel). 2024; 17(4).

PMID: 38399077 PMC: 10890344. DOI: 10.3390/ma17040827.


Effect of Stabilized Martensite on the Long-Term Performance of Superelastic NiTi Endodontic Files.

Sanchez P, Vidi B, Mena-Alvarez J, Gil J, Rico C, Aragoneses J Materials (Basel). 2023; 16(11).

PMID: 37297223 PMC: 10254520. DOI: 10.3390/ma16114089.


3D Surface Scanning-A Novel Protocol to Characterize Virtual Nickel-Titanium Endodontic Instruments.

Martins J, Pinto R, Silva E, Simoes-Carvalho M, Marques D, Martins R Materials (Basel). 2023; 16(10).

PMID: 37241263 PMC: 10222178. DOI: 10.3390/ma16103636.


Cyclic fatigue resistance of EdgeTaper Platinum, Protaper Gold, and TruNatomy Prime rotary files before and after autoclave sterilization.

Almohareb R, Barakat R, Algahtani F, Alkadi M PeerJ. 2023; 11:e14656.

PMID: 36700004 PMC: 9869776. DOI: 10.7717/peerj.14656.


Effect of Rotational Speed on the Resistance of NiTi Alloy Endodontic Rotary Files to Cyclic Fatigue-An In Vitro Study.

Faus-Matoses V, Faus-Llacer V, Ruiz-Sanchez C, Jaramillo-Vasconez S, Faus-Matoses I, Martin-Biedma B J Clin Med. 2022; 11(11).

PMID: 35683530 PMC: 9181116. DOI: 10.3390/jcm11113143.


References
1.
Kwak S, Ha J, Lee C, Abed R, Abu-Tahun I, Kim H . Effects of Pitch Length and Heat Treatment on the Mechanical Properties of the Glide Path Preparation Instruments. J Endod. 2016; 42(5):788-92. DOI: 10.1016/j.joen.2016.02.002. View

2.
Schneider S . A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol. 1971; 32(2):271-5. DOI: 10.1016/0030-4220(71)90230-1. View

3.
Sjogren U, Figdor D, Persson S, Sundqvist G . Influence of infection at the time of root filling on the outcome of endodontic treatment of teeth with apical periodontitis. Int Endod J. 1998; 30(5):297-306. DOI: 10.1046/j.1365-2591.1997.00092.x. View

4.
Uygun A, Unal M, Falakaloglu S, Guven Y . Comparison of the cyclic fatigue resistance of hyflex EDM, vortex blue, protaper gold, and onecurve nickel-Titanium instruments. Niger J Clin Pract. 2020; 23(1):41-45. DOI: 10.4103/njcp.njcp_343_19. View

5.
Turpin Y, Chagneau F, Vulcain J . Impact of two theoretical cross-sections on torsional and bending stresses of nickel-titanium root canal instrument models. J Endod. 2001; 26(7):414-7. DOI: 10.1097/00004770-200007000-00009. View