Simulating the Time Evolving Geometry, Mechanical Properties, and Fibrous Structure of Bioprosthetic Heart Valve Leaflets Under Cyclic Loading

Overview
Publisher Elsevier
Date 2021 Sep 5
PMID 34482092
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Currently, the most common replacement heart valve design is the 'bioprosthetic' heart valve (BHV), which has important advantages in that it does not require permanent anti-coagulation therapy, operates noiselessly, and has blood flow characteristics similar to the native valve. BHVs are typically fabricated from glutaraldehyde-crosslinked pericardial xenograft tissue biomaterials (XTBs) attached to a rigid, semi-flexible, or fully collapsible stent in the case of the increasingly popular transcutaneous aortic valve replacement (TAVR). While current TAVR assessments are positive, clinical results to date are generally limited to <2 years. Since TAVR leaflets are constructed using thinner XTBs, their mechanical demands are substantially greater than surgical BHV due to the increased stresses during in vivo operation, potentially resulting in decreased durability. Given the functional complexity of heart valve operation, in-silico predictive simulations clearly have potential to greatly improve the TAVR development process. As such simulations must start with accurate material models, we have developed a novel time-evolving constitutive model for pericardial xenograft tissue biomaterials (XTB) utilized in BHV (doi: 10.1016/j.jmbbm.2017.07.013). This model was able to simulate the observed tissue plasticity effects that occur in approximately in the first two years of in vivo function (50 million cycles). In the present work, we implemented this model into a complete simulation pipeline to predict the BHV time evolving geometry to 50 million cycles. The pipeline was implemented within an isogeometric finite element formulation that directly integrated our established BHV NURBS-based geometry (doi: 10.1007/s00466-015-1166-x). Simulations of successive loading cycles indicated continual changes in leaflet shape, as indicated by spatially varying increases in leaflet curvature. While the simulation model assumed an initial uniform fiber orientation distribution, anisotropic regional changes in leaflet tissue plastic strain induced a complex changes in regional fiber orientation. We have previously noted in our time-evolving constitutive model that the increases in collagen fiber recruitment with cyclic loading placed an upper bound on plastic strain levels. This effect was manifested by restricting further changes in leaflet geometry past 50 million cycles. Such phenomena was accurately captured in the valve-level simulations due to the use of a tissue-level structural-based modeling approach. Changes in basic leaflet dimensions agreed well with extant experimental studies. As a whole, the results of the present study indicate the complexity of BHV responses to cyclic loading, including changes in leaflet shape and internal fibrous structure. It should be noted that the later effect also influences changes in local mechanical behavior (i.e. changes in leaflet anisotropic tissue stress-strain relationship) due to internal fibrous structure resulting from plastic strains. Such mechanism-based simulations can help pave the way towards the application of sophisticated simulation technologies in the development of replacement heart valve technology.

Citing Articles

Computational analysis of heart valve growth and remodeling after the Ross procedure.

Middendorp E, Braeu F, Baaijens F, Humphrey J, Cyron C, Loerakker S Biomech Model Mechanobiol. 2024; 23(6):1889-1907.

PMID: 39269523 PMC: 11554944. DOI: 10.1007/s10237-024-01874-y.


Identifying Heterogeneous Micromechanical Properties of Biological Tissues via Physics-Informed Neural Networks.

Wu W, Daneker M, Turner K, Jolley M, Lu L Small Methods. 2024; 9(1):e2400620.

PMID: 39091065 PMC: 11747890. DOI: 10.1002/smtd.202400620.


Identifying heterogeneous micromechanical properties of biological tissues via physics-informed neural networks.

Wu W, Daneker M, Turner K, Jolley M, Lu L ArXiv. 2024; .

PMID: 38745694 PMC: 11092874.


Current progress toward isogeometric modeling of the heart biophysics.

Torre M, Morganti S, Pasqualini F, Reali A Biophys Rev (Melville). 2024; 4(4):041301.

PMID: 38510845 PMC: 10903424. DOI: 10.1063/5.0152690.


Leaflet Mechanical Stress in Different Designs and Generations of Transcatheter Aortic Valves: An in Vitro Study.

Stanova V, Rieu R, Thollon L, Salaun E, Rodes-Cabau J, Cote N Struct Heart. 2024; 8(2):100262.

PMID: 38481716 PMC: 10927450. DOI: 10.1016/j.shj.2023.100262.


References
1.
Martin C, Sun W . Modeling of long-term fatigue damage of soft tissue with stress softening and permanent set effects. Biomech Model Mechanobiol. 2012; 12(4):645-55. PMC: 3546129. DOI: 10.1007/s10237-012-0431-6. View

2.
Thubrikar M, DECK J, Aouad J, Nolan S . Role of mechanical stress in calcification of aortic bioprosthetic valves. J Thorac Cardiovasc Surg. 1983; 86(1):115-25. View

3.
Lanir Y . Constitutive equations for fibrous connective tissues. J Biomech. 1983; 16(1):1-12. DOI: 10.1016/0021-9290(83)90041-6. View

4.
Kamensky D, Hsu M, Schillinger D, Evans J, Aggarwal A, Bazilevs Y . An immersogeometric variational framework for fluid-structure interaction: application to bioprosthetic heart valves. Comput Methods Appl Mech Eng. 2014; 284:1005-1053. PMC: 4274080. DOI: 10.1016/j.cma.2014.10.040. View

5.
Pomar J, Bosch X, Chaitman B, Pelletier C, Grondin C . Late tears in leaflets of porcine bioprostheses in adults. Ann Thorac Surg. 1984; 37(1):78-83. DOI: 10.1016/s0003-4975(10)60716-3. View