» Articles » PMID: 34441856

Comparison of Histological Sample Volumes Among Various Endoscopic Ultrasound-Guided Biopsy Needles

Overview
Journal J Clin Med
Specialty General Medicine
Date 2021 Aug 27
PMID 34441856
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Endoscopic ultrasound-guided fine-needle biopsy (EUS-FNB) enables easy and accurate pathological assessment. Here, we compared and assessed the area of samples on glass slides for three needle types: a 19-gauge Franseen needle (Acquire, Boston Scientific, Natick, MA, USA), a 22-gauge Franseen needle, and a 19-gauge fine-needle aspiration (FNA) needle (EZ Shot 3 Plus; Olympus, Tokyo, Japan). Among patients with suspected pancreatic cancer, with a ≥20 mm tumor located in the pancreatic body and tail, and who underwent EUS-FNA or FNB between June 2018 and March 2020, 10 were randomly selected to test each needle. The areas of histological tissue and blood clot samples were measured using the BZ-X800 imaging software (Keyence Corporation, Osaka, Japan). Baseline patient characteristics and pathological sample data showed no significant differences among the needles. The 19-gauge Franseen needle obtained significantly more histological tissue samples than the 19-gauge conventional needle ( = 0.010) and 22-gauge Franseen needle ( = 0.008). Conversely, there was no significant difference between the 19-gauge conventional needle and 22-gauge Franseen needle ( = 0.838) in this regard. The 19-gauge Franseen needle could collect more samples than the other needles, contributing to giving a more precise pathological diagnosis and more information, including genomic profiling.

Citing Articles

Clinical utility of endoscopic ultrasound-guided tissue acquisition for comprehensive genomic profiling of patients with biliary tract cancer, especially with intrahepatic cholangiocarcinoma.

Yanaidani T, Hara K, Okuno N, Haba S, Kuwahara T, Kuraishi Y Clin Endosc. 2024; 57(3):384-392.

PMID: 38356172 PMC: 11133989. DOI: 10.5946/ce.2023.139.


Cutting edge of endoscopic ultrasound-guided fine-needle aspiration for solid pancreatic lesions.

Ishikawa T, Yamao K, Mizutani Y, Iida T, Kawashima H J Med Ultrason (2001). 2023; 51(2):209-217.

PMID: 37914883 PMC: 11098899. DOI: 10.1007/s10396-023-01375-y.


Comparison of the novel Franseen needle versus the fine-needle aspiration needle in endoscopic ultrasound-guided tissue acquisition for cancer gene panel testing: A propensity score-matching analysis.

Mori T, Ozawa E, Shimakura A, Takahashi K, Matsuo S, Tajima K JGH Open. 2023; 7(9):652-658.

PMID: 37744704 PMC: 10517436. DOI: 10.1002/jgh3.12965.


Current status and issues in genomic analysis using EUS-FNA/FNB specimens in hepatobiliary-pancreatic cancers.

Ozono Y, Kawakami H, Uchiyama N, Hatada H, Ogawa S J Gastroenterol. 2023; 58(11):1081-1093.

PMID: 37698719 PMC: 10590314. DOI: 10.1007/s00535-023-02037-z.


Diagnostic Ability of Endoscopic Ultrasound-Guided Tissue Acquisition Using 19-Gauge Fine-Needle Biopsy Needle for Abdominal Lesions.

Takeshita K, Hijioka S, Nagashio Y, Maruki Y, Kawasaki Y, Maehara K Diagnostics (Basel). 2023; 13(3).

PMID: 36766558 PMC: 9914510. DOI: 10.3390/diagnostics13030450.


References
1.
Kamisawa T, Wood L, Itoi T, Takaori K . Pancreatic cancer. Lancet. 2016; 388(10039):73-85. DOI: 10.1016/S0140-6736(16)00141-0. View

2.
Polkowski M, Larghi A, Weynand B, Boustiere C, Giovannini M, Pujol B . Learning, techniques, and complications of endoscopic ultrasound (EUS)-guided sampling in gastroenterology: European Society of Gastrointestinal Endoscopy (ESGE) Technical Guideline. Endoscopy. 2011; 44(2):190-206. DOI: 10.1055/s-0031-1291543. View

3.
Mita N, Iwashita T, Uemura S, Iwasa Y, Toda K, Mukai T . Endoscopic Ultrasound-Guided Fine Needle Biopsy Using 22-Gauge Franseen Needle for the Histological Diagnosis of Solid Lesions: A Multicenter Prospective Pilot Study. Dig Dis Sci. 2019; 65(4):1155-1163. DOI: 10.1007/s10620-019-05840-y. View

4.
Adler D, Muthusamy V, Ehrlich D, Parasher G, Thosani N, Chen A . A multicenter evaluation of a new EUS core biopsy needle: Experience in 200 patients. Endosc Ultrasound. 2018; 8(2):99-104. PMC: 6482605. DOI: 10.4103/eus.eus_53_17. View

5.
Bang J, Hebert-Magee S, Hasan M, Navaneethan U, Hawes R, Varadarajulu S . Endoscopic ultrasonography-guided biopsy using a Franseen needle design: Initial assessment. Dig Endosc. 2016; 29(3):338-346. DOI: 10.1111/den.12769. View