A Review of the Literature on the Accuracy, Strengths, and Limitations of Visual, Thoracic Impedance, and Electrocardiographic Methods Used to Measure Respiratory Rate in Hospitalized Patients

Overview
Date 2021 Aug 18
PMID 34405488
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Respiratory rate (RR) is one of the most important indicators of a patient's health. In critically ill patients, unrecognized changes in RR are associated with poorer outcomes. Visual assessment (VA), impedance pneumography (IP), and electrocardiographic-derived respiration (EDR) are the three most commonly used methods to assess RR. While VA and IP are widely used in hospitals, the EDR method has not been validated for use in hospitalized patients. Additionally, little is known about their accuracy compared with one another. The purpose of this systematic review was to compare the accuracy, strengths, and limitations of VA of RR to two methods that use physiologic data, namely IP and EDR.

Methods: A systematic review of the literature was undertaken using prespecified inclusion and exclusion criteria. Each of the studies was evaluated using standardized criteria.

Results: Full manuscripts for 23 studies were reviewed, and four studies were included in this review. Three studies compared VA to IP and one study compared VA to EDR. In terms of accuracy, when Bland-Altman analyses were performed, the upper and lower levels of agreement were extremely poor for both the VA and IP and VA and EDR comparisons.

Conclusion: Given the paucity of research and the fact that no studies have compared all three methods, no definitive conclusions can be drawn about the accuracy of these three methods. The clinical importance of accurate assessment of RR warrants new research with rigorous designs to determine the accuracy, and clinically meaningful levels of agreement of these methods.

Citing Articles

Contactless and continuous monitoring of respiratory rate in a hospital ward: a clinical validation study.

Toften S, Kjellstadli J, Kvaerness J, Pedersen L, Laugsand L, Thu O Front Physiol. 2024; 15:1502413.

PMID: 39665054 PMC: 11631942. DOI: 10.3389/fphys.2024.1502413.


Day-to-Day Variability in Measurements of Respiration Using Bioimpedance from a Non-Standard Location.

Goyal K, Shah D, Day S Sensors (Basel). 2024; 24(14).

PMID: 39066010 PMC: 11280867. DOI: 10.3390/s24144612.


Methodological quality of 100 recent systematic reviews of health-related outcome measurement instruments: an overview of reviews.

Elsman E, Mokkink L, Abma I, Aiyegbusi O, Chiarotto A, Haywood K Qual Life Res. 2024; 33(10):2593-2609.

PMID: 38961010 PMC: 11452433. DOI: 10.1007/s11136-024-03706-z.


Respiratory rate measurement by pressure variation in the high flow nasal cannula-system in healthy volunteers.

Miechels J, Koning M J Clin Monit Comput. 2024; 38(6):1397-1404.

PMID: 38867018 DOI: 10.1007/s10877-024-01185-8.


Laser spectroscopic method for remote sensing of respiratory rate.

Bachir W, Ismael F, Alaineya N Phys Eng Sci Med. 2023; 46(3):1249-1258.

PMID: 37358781 PMC: 10480269. DOI: 10.1007/s13246-023-01292-x.


References
1.
Helfenbein E, Firoozabadi R, Chien S, Carlson E, Babaeizadeh S . Development of three methods for extracting respiration from the surface ECG: a review. J Electrocardiol. 2014; 47(6):819-25. DOI: 10.1016/j.jelectrocard.2014.07.020. View

2.
Lovett P, Buchwald J, Sturmann K, Bijur P . The vexatious vital: neither clinical measurements by nurses nor an electronic monitor provides accurate measurements of respiratory rate in triage. Ann Emerg Med. 2005; 45(1):68-76. DOI: 10.1016/j.annemergmed.2004.06.016. View

3.
Cooper S, Cant R, Sparkes L . Respiratory rate records: the repeated rate?. J Clin Nurs. 2013; 23(9-10):1236-8. DOI: 10.1111/jocn.12234. View

4.
Kamio T, Kajiwara A, Iizuka Y, Shiotsuka J, Sanui M . Frequency of vital sign measurement among intubated patients in the general ward and nurses' attitudes toward vital sign measurement. J Multidiscip Healthc. 2018; 11:575-581. PMC: 6197211. DOI: 10.2147/JMDH.S179033. View

5.
Varon C, Morales J, Lazaro J, Orini M, Deviaene M, Kontaxis S . A Comparative Study of ECG-derived Respiration in Ambulatory Monitoring using the Single-lead ECG. Sci Rep. 2020; 10(1):5704. PMC: 7109157. DOI: 10.1038/s41598-020-62624-5. View