» Articles » PMID: 33996213

Clinical Integration of Fast Raman Spectroscopy for Mohs Micrographic Surgery of Basal Cell Carcinoma

Overview
Specialty Radiology
Date 2021 May 17
PMID 33996213
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

We present the first clinical integration of a prototype device based on integrated auto-fluorescence imaging and Raman spectroscopy (Fast Raman device) for intra-operative assessment of surgical margins during Mohs micrographic surgery of basal cell carcinoma (BCC). Fresh skin specimens from 112 patients were used to optimise the tissue pre-processing and the Fast Raman algorithms to enable an analysis of complete Mohs layers within 30 minutes. The optimisation allowed >95% of the resection surface area to be investigated (including the deep and epidermal margins). The Fast Raman device was then used to analyse skin layers excised from the most relevant anatomical sites (nose, temple, eyelid, cheek, forehead, eyebrow and lip) and to detect the three main types of BCC (nodular, superficial and infiltrative). These results suggest that the Fast Raman technique is a promising tool to provide an objective diagnosis "tumour clear yes/no" during Mohs surgery of BCC. This clinical integration study is a key step towards a larger scale diagnosis test accuracy study to reliably determine the sensitivity and specificity in a clinical setting.

Citing Articles

Application of the OSNA Technique (One-Step Nucleic Acid Amplification Test) in Breast Cancer.

Sychowski G, Romanowicz H, Smolarz B Int J Mol Sci. 2025; 26(2).

PMID: 39859370 PMC: 11766269. DOI: 10.3390/ijms26020656.


Co-localized line-field confocal optical coherence tomography and confocal Raman microspectroscopy for three-dimensional high-resolution morphological and molecular characterization of skin tissues .

Waszczuk L, Ogien J, Perrot J, Dubois A Biomed Opt Express. 2022; 13(4):2467-2487.

PMID: 35519243 PMC: 9045904. DOI: 10.1364/BOE.450993.


The complementary value of intraoperative fluorescence imaging and Raman spectroscopy for cancer surgery: combining the incompatibles.

Lauwerends L, Abbasi H, Bakker Schut T, Van Driel P, Hardillo J, Santos I Eur J Nucl Med Mol Imaging. 2022; 49(7):2364-2376.

PMID: 35102436 PMC: 9165240. DOI: 10.1007/s00259-022-05705-z.


Emerging Minimally Invasive Technologies for the Detection of Skin Cancer.

Jung J, Cho J, Lee W, Chang S, Lee M, Won C J Pers Med. 2021; 11(10).

PMID: 34683091 PMC: 8538732. DOI: 10.3390/jpm11100951.

References
1.
Longo C, Pampena R, Bombonato C, Gardini S, Piana S, Mirra M . Diagnostic accuracy of ex vivo fluorescence confocal microscopy in Mohs surgery of basal cell carcinomas: a prospective study on 753 margins. Br J Dermatol. 2018; 180(6):1473-1480. DOI: 10.1111/bjd.17507. View

2.
Takamori S, Kong K, Varma S, Leach I, Williams H, Notingher I . Optimization of multimodal spectral imaging for assessment of resection margins during Mohs micrographic surgery for basal cell carcinoma. Biomed Opt Express. 2015; 6(1):98-111. PMC: 4317116. DOI: 10.1364/BOE.6.000098. View

3.
Baxter J, Patel A, Varma S . Facial basal cell carcinoma. BMJ. 2012; 345:e5342. DOI: 10.1136/bmj.e5342. View

4.
Bastiaens M, Hoefnagel J, Bruijn J, Westendorp R, Vermeer B, Bouwes Bavinck J . Differences in age, site distribution, and sex between nodular and superficial basal cell carcinoma indicate different types of tumors. J Invest Dermatol. 1998; 110(6):880-4. DOI: 10.1046/j.1523-1747.1998.00217.x. View

5.
Durkin J, Fine J, Sam H, Pugliano-Mauro M, Ho J . Imaging of Mohs micrographic surgery sections using full-field optical coherence tomography: a pilot study. Dermatol Surg. 2014; 40(3):266-74. DOI: 10.1111/dsu.12419. View