Elliptical Pipette Generated Large Microdroplets for POC Visual DdPCR Quantification of Low Viral Load
Authors
Affiliations
Rapid point-of-care (POC) quantification of low virus RNA load would significantly reduce the turn-around time for the PCR test and help contain a fast-spreading epidemic. Herein, we report a droplet digital PCR (ddPCR) platform that can achieve this sensitivity and rapidity without bulky lab-bound equipment. The key technology is a flattened pipette tip with an elliptical cross-section, which extends a high aspect-ratio microfluidic chip design to pipette scale, for rapid (<5 min) generation of several thousand monodispersed droplets ∼150 to 350 μm in size with a CV of ∼2.3%. A block copolymer surfactant (polyoxyalkylene F127) is used to stabilize these large droplets in oil during thermal cycling. At this droplet size and number, positive droplets can be counted by eye or imaged by a smartphone with appropriate illumination/filtering to accurately quantify up to 100 target copies. We demonstrate with 2019 nCoV-PCR assay LODs of 3.8 copies per 20 μL of sample and a dynamic range of 4-100 copies. The ddPCR platform is shown to be inhibitor resistant with spiked saliva samples, suggesting RNA extraction may not be necessary. It represents a rapid 1.5-h POC quantitative PCR test that requires just a pipette equipped with elliptical pipette tip, a commercial portable thermal cycler, a smartphone, and a portable trans-illuminator, without bulky and expensive micropumps and optical detectors that prevent POC application.
Interfacial tension driven open droplet microfluidics.
Khor J, Lee U, Berthier J, Berthier E, Theberge A Adv Mater Interfaces. 2024; 10(7).
PMID: 39584054 PMC: 11583357. DOI: 10.1002/admi.202202234.
Advances in Nucleic Acid Assays for Infectious Disease: The Role of Microfluidic Technology.
Wang Y, Chen J, Yang Z, Wang X, Zhang Y, Chen M Molecules. 2024; 29(11).
PMID: 38893293 PMC: 11173870. DOI: 10.3390/molecules29112417.
[Application of Droplet-Based Microfluidics in Microbial Research].
Li Z, Peng X Sichuan Da Xue Xue Bao Yi Xue Ban. 2023; 54(3):673-678.
PMID: 37248604 PMC: 10475413. DOI: 10.12182/20230560303.
Application of microfluidic technologies on COVID-19 diagnosis and drug discovery.
Lin Z, Zou Z, Pu Z, Wu M, Zhang Y Acta Pharm Sin B. 2023; .
PMID: 36855672 PMC: 9951611. DOI: 10.1016/j.apsb.2023.02.014.
Cost-Effective Droplet Generator for Portable Bio-Applications.
Du L, Li Y, Wang J, Zhou Z, Lan T, Jing D Micromachines (Basel). 2023; 14(2).
PMID: 36838166 PMC: 9966110. DOI: 10.3390/mi14020466.