A Multi-inversion Multi-echo Spin and Gradient Echo Echo Planar Imaging Sequence with Low Image Distortion for Rapid Quantitative Parameter Mapping and Synthetic Image Contrasts

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2021 Mar 25
PMID 33764563
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Brain imaging exams typically take 10-20 min and involve multiple sequential acquisitions. A low-distortion whole-brain echo planar imaging (EPI)-based approach was developed to efficiently encode multiple contrasts in one acquisition, allowing for calculation of quantitative parameter maps and synthetic contrast-weighted images.

Methods: Inversion prepared spin- and gradient-echo EPI was developed with slice-order shuffling across measurements for efficient acquisition with T , T , and weighting. A dictionary-matching approach was used to fit the images to quantitative parameter maps, which in turn were used to create synthetic weighted images with typical clinical contrasts. Dynamic slice-optimized multi-coil shimming with a B shim array was used to reduce B inhomogeneity and, therefore, image distortion by >50%. Multi-shot EPI was also implemented to minimize distortion and blurring while enabling high in-plane resolution. A low-rank reconstruction approach was used to mitigate errors from shot-to-shot phase variation.

Results: The slice-optimized shimming approach was combined with in-plane parallel-imaging acceleration of 4× to enable single-shot EPI with more than eight-fold distortion reduction. The proposed sequence efficiently obtained 40 contrasts across the whole-brain in just over 1 min at 1.2 × 1.2 × 3 mm resolution. The multi-shot variant of the sequence achieved higher in-plane resolution of 1 × 1 × 4 mm with good image quality in 4 min. Derived quantitative maps showed comparable values to conventional mapping methods.

Conclusion: The approach allows fast whole-brain imaging with quantitative parameter maps and synthetic weighted contrasts. The slice-optimized multi-coil shimming and multi-shot reconstruction approaches result in minimal EPI distortion, giving the sequence the potential to be used in rapid screening applications.

Citing Articles

Synthetic MR: Clinical applications in neuroradiology.

Sharma S, Nayak A, Thomas B, Kesavadas C Neuroradiology. 2025; .

PMID: 39888426 DOI: 10.1007/s00234-025-03547-8.


A single 1-min brain MRI scan for generating multiple synthetic image contrasts in awake children from quantitative relaxometry maps.

Kilpattu Ramaniharan A, Pednekar A, Parikh N, Nagaraj U, Manhard M Pediatr Radiol. 2024; 55(2):312-323.

PMID: 39692886 DOI: 10.1007/s00247-024-06113-1.


Optimization and validation of multi-echo, multi-contrast SAGE acquisition in fMRI.

Keeling E, Bergamino M, Ragunathan S, Quarles C, Newton A, Stokes A Imaging Neurosci (Camb). 2024; 2:1-20.

PMID: 39449748 PMC: 11497078. DOI: 10.1162/imag_a_00217.


Simultaneous Multiparameter Mapping of the Liver in a Single Breath-Hold or Respiratory-Triggered Acquisition Using Multi-Inversion Spin and Gradient Echo MRI.

Manhard M, Kilpattu Ramaniharan A, Tkach J, Trout A, Dillman J, Pednekar A J Magn Reson Imaging. 2024; 61(4):1925-1936.

PMID: 39192381 PMC: 11896940. DOI: 10.1002/jmri.29584.


Multidimensional MRI for Characterization of Subtle Axonal Injury Accelerated Using an Adaptive Nonlocal Multispectral Filter.

Benjamini D, Bouhrara M, Komlosh M, Iacono D, Perl D, Brody D Front Phys. 2023; 9.

PMID: 37408700 PMC: 10321473. DOI: 10.3389/fphy.2021.737374.


References
1.
Eichner C, Cauley S, Cohen-Adad J, Moller H, Turner R, Setsompop K . Real diffusion-weighted MRI enabling true signal averaging and increased diffusion contrast. Neuroimage. 2015; 122:373-84. PMC: 4651971. DOI: 10.1016/j.neuroimage.2015.07.074. View

2.
Feinberg D, Moeller S, Smith S, Auerbach E, Ramanna S, Gunther M . Multiplexed echo planar imaging for sub-second whole brain FMRI and fast diffusion imaging. PLoS One. 2010; 5(12):e15710. PMC: 3004955. DOI: 10.1371/journal.pone.0015710. View

3.
Schmiedeskamp H, Straka M, Newbould R, Zaharchuk G, Andre J, Olivot J . Combined spin- and gradient-echo perfusion-weighted imaging. Magn Reson Med. 2011; 68(1):30-40. PMC: 3374915. DOI: 10.1002/mrm.23195. View

4.
Wang D, Ostenson J, Smith D . snapMRF: GPU-accelerated magnetic resonance fingerprinting dictionary generation and matching using extended phase graphs. Magn Reson Imaging. 2019; 66:248-256. PMC: 7031031. DOI: 10.1016/j.mri.2019.11.015. View

5.
Setsompop K, Gagoski B, Polimeni J, Witzel T, Wedeen V, Wald L . Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty. Magn Reson Med. 2011; 67(5):1210-24. PMC: 3323676. DOI: 10.1002/mrm.23097. View