Comparison Between Two Devices in the Detection of Corneal Thickness Changes After Cataract Surgery

Overview
Journal Sci Rep
Specialty Science
Date 2021 Mar 24
PMID 33758298
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

This study compared corneal thickness (CT) changes obtained with specular microscopy (SM) and a rotating Scheimpflug camera (RSC) after conventional phacoemulsification surgery (PS). One hundred sixty six eyes of 83 patients were analyzed before and one month after PS. One eye underwent PS, while the fellow phakic one was used as control. CT was measured with SM at the center of the cornea and with RSC at the pupil center, at the corneal apex and at the thinnest point. In the operated eye, SM showed a larger CT mean increase than those one detected at the three different measurements' points evaluated by RSC. Inversely, in the fellow phakic eye, SM showed a greater CT mean decrease than those one registered by RSC at its three measurement's points. Thus, one month after surgery, even if cornea appears clear at the slit-lamp, a significant thickness increase is still present. This is even more evident if the slight decrease of the fellow phakic eye is considered. The differences between the two devices are probably related to the different measured areas.

Citing Articles

Intraocular Pressure Measurements in Standing, Sitting, Prone, and Supine Positions.

De Bernardo M, Cione F, De Pascale I, Pagliarulo S, Rosa N J Pers Med. 2024; 14(8).

PMID: 39202017 PMC: 11355470. DOI: 10.3390/jpm14080826.


Intraocular Pressure Measurements in Standing, Sitting, and Supine Position: Comparison between Tono-Pen Avia and Icare Pro Tonometers.

De Bernardo M, Abbinante G, Borrelli M, Di Stasi M, Cione F, Rosa N J Clin Med. 2022; 11(21).

PMID: 36362460 PMC: 9654527. DOI: 10.3390/jcm11216234.


Recent Evidence of Tea Tree Oil Effectiveness in Blepharitis Treatment.

Capasso L, Abbinante G, Coppola A, Salerno G, De Bernardo M Biomed Res Int. 2022; 2022:9204251.

PMID: 35941981 PMC: 9356889. DOI: 10.1155/2022/9204251.


Evaluation of Corneal Structure and Endothelial Morphological Characteristics in Type 2 Diabetic and Non-Diabetic Patients [Letter].

Graziano M, Capasso L, Rosa N Clin Ophthalmol. 2021; 15:3539-3540.

PMID: 34465973 PMC: 8397702. DOI: 10.2147/OPTH.S332774.


Anterior Chamber Depth After Phacovitrectomy [Letter].

De Bernardo M, Di Paola I, Rosa N Clin Ophthalmol. 2021; 15:2741-2742.

PMID: 34211262 PMC: 8240845. DOI: 10.2147/OPTH.S321905.

References
1.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

2.
De Bernardo M, Rosa N . Central Corneal Thickness After Cross-linking Using High-Definition Optical Coherence Tomography, Ultrasound, and Dual Scheimpflug Tomography: A Comparative Study Over One Year. Am J Ophthalmol. 2017; 176:254. DOI: 10.1016/j.ajo.2017.01.007. View

3.
Jonuscheit S, Doughty M . Regional repeatability measures of corneal thickness: Orbscan II and ultrasound. Optom Vis Sci. 2007; 84(1):52-8. DOI: 10.1097/01.opx.0000254045.62252.b4. View

4.
Sarnicola C, Sarnicola E, Panico E, Panico C, Sarnicola V . Cataract surgery in corneal transplantation. Curr Opin Ophthalmol. 2019; 31(1):23-27. DOI: 10.1097/ICU.0000000000000635. View

5.
de Juan V, Herreras J, Perez I, Morejon A, Rio-Cristobal A, Rio-San Cristobal A . Refractive stabilization and corneal swelling after cataract surgery. Optom Vis Sci. 2012; 90(1):31-6. DOI: 10.1097/OPX.0b013e318278fc44. View