» Articles » PMID: 33684584

Use of the Scan-and-Plan Workflow in Next-Generation Robot-Assisted Pedicle Screw Insertion: Retrospective Cohort Study and Literature Review

Overview
Journal World Neurosurg
Publisher Elsevier
Date 2021 Mar 8
PMID 33684584
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To report our experience using the scan-and-plan workflow and review current literature on surgical efficiency, safety, and accuracy of next-generation robot-assisted (RA) spine surgery.

Methods: The records of patients who underwent RA pedicle screw fixation were reviewed. The accuracy of pedicle screw placement was determined based on the Ravi classification system. To evaluate workflow efficiency, 3 demographically matched cohorts were created to analyze differences in time per screw placement (defined as operating room [OR] time divided by number of screws placed). Group A had <4 screws placed, Group B had 4 screws placed, and Group C had >4 screws placed. Intraoperative errors and postoperative complications were collected to elucidate safety.

Results: Eighty-four RA cases (306 pedicle screws) were included for analysis. The mean number of screws placed was 2.1 ± 0.3 in Group A and 6.4 ± 1.2 in Group C; 4 screws were placed in Group B patients. The accuracy rate (Ravi grade I) was 98.4%. Screw placement time was significantly longer in Group A (101 ± 37.7 minutes) than Group B (50.5 ± 25.4 minutes) or C (43.6 ± 14.7 minutes). There were no intraoperative complications, robot failures, or in-hospital complications requiring a return to the OR.

Conclusions: The scan-and-plan workflow allowed for a high degree of accuracy. It was a safe method that provided a smooth and efficient OR workflow without registration errors or robotic failures. After the placement of 4 pedicle screws, the per-screw time remained constant. Further studies regarding efficiency and utility in multilevel procedures are necessary.

Citing Articles

Flattening the learning curve - Early experience of robotic-assisted pedicle screw placement in spine surgery.

Srinivasa V, Thirugnanam B, Kanhangad M, Soni A, Kashyap A, Vidyadhara A J Orthop. 2024; 57:49-54.

PMID: 38973970 PMC: 11225720. DOI: 10.1016/j.jor.2024.06.011.


Comparison of No Tap (two-step) and tapping robotic assisted cortical bone trajectory screw insertion.

Werthmann 3rd N, Gum J, Nagata K, Djurasovic M, Glassman S, Owens 2nd R J Robot Surg. 2024; 18(1):204.

PMID: 38714574 DOI: 10.1007/s11701-024-01890-1.