Objective: To evaluate the trueness and precision of full-arch scans acquired using five intraoral scanners and investigate the factors associated with the dimensional accuracy of the intraoral scan data.

Methods: Nine adult participants (mean age, 34.3 ± 8.3 years) were recruited. Four zirconium spheres (Ø 6 mm) were bonded to the canines and the molars. Following acquisition of reference scans using an industrial-grade scanner, five intraoral scanners, namely i500, CS3600, Trios 3, iTero, and CEREC Omnicam, were used to scan the arches. Linear distances between the four reference spheres were automatically calculated, and linear mixed model analysis was performed to compare the trueness and precision of the intraoral scan data among the different scanners.

Results: The absolute mean trueness and precision values for all intraoral scanners were 76.6 ± 79.3 and 56.6 ± 52.4 μm, respectively. The type of scanner and the measured linear distances had significant effects on the accuracy of the intraoral scan data. With regard to trueness, errors in the intermolar dimension and the distance from the canine to the contralateral molar were greater with Omnicam than with the other scanners. With regard to precision, the error in the linear distance from the canine to the molar in the same quadrant was greater with Omnicam and CS3600 than with the other scanners.

Conclusions: The dimensional accuracy of intraoral scan data may differ significantly according to the type of scanner, with the amount of error in terms of trueness being clinically significant.

Citing Articles

Comparisons of precision and trueness of digital dental casts produced by desktop scanners and intraoral scanners.

Chuang C, Cheng F, Chen M, Lin W, Chiang C J Dent Sci. 2025; 20(1):137-146.

PMID: 39873032 PMC: 11763548. DOI: 10.1016/j.jds.2024.09.016.


Evaluation of the accuracy of full-arch impressions between three different intraoral scanners and conventional impressions: A prospective in vivo study.

Bhatia N, Pandian S J Dent Res Dent Clin Dent Prospects. 2024; 18(1):77-84.

PMID: 38881640 PMC: 11179144. DOI: 10.34172/joddd.40841.


Surface wear of attachments in patients during clear aligner therapy: a prospective clinical study.

Li Q, Yang K Prog Orthod. 2024; 25(1):7.

PMID: 38369617 PMC: 10874919. DOI: 10.1186/s40510-023-00506-y.


Accuracy of Computerized Optical Impression Making in Fabrication of Removable Dentures for Partially Edentulous Jaws: An In Vivo Feasibility Study.

Saravi B, Ilbertz J, Vach K, Kohal R, Patzelt S J Funct Biomater. 2023; 14(9).

PMID: 37754872 PMC: 10532381. DOI: 10.3390/jfb14090458.


Intraoral scanners in children: evaluation of the patient perception, reliability and reproducibility, and chairside time-A systematic review.

Serrano-Velasco D, Martin-Vacas A, Paz-Cortes M, Giovannini G, Cintora-Lopez P, Aragoneses J Front Pediatr. 2023; 11:1213072.

PMID: 37435173 PMC: 10331299. DOI: 10.3389/fped.2023.1213072.


References
1.
Goracci C, Franchi L, Vichi A, Ferrari M . Accuracy, reliability, and efficiency of intraoral scanners for full-arch impressions: a systematic review of the clinical evidence. Eur J Orthod. 2015; 38(4):422-8. DOI: 10.1093/ejo/cjv077. View

2.
Ender A, Attin T, Mehl A . In vivo precision of conventional and digital methods of obtaining complete-arch dental impressions. J Prosthet Dent. 2015; 115(3):313-20. DOI: 10.1016/j.prosdent.2015.09.011. View

3.
Boeddinghaus M, Breloer E, Rehmann P, Wostmann B . Accuracy of single-tooth restorations based on intraoral digital and conventional impressions in patients. Clin Oral Investig. 2015; 19(8):2027-34. DOI: 10.1007/s00784-015-1430-7. View

4.
Li H, Lyu P, Wang Y, Sun Y . Influence of object translucency on the scanning accuracy of a powder-free intraoral scanner: A laboratory study. J Prosthet Dent. 2016; 117(1):93-101. DOI: 10.1016/j.prosdent.2016.04.008. View

5.
Flugge T, Schlager S, Nelson K, Nahles S, Metzger M . Precision of intraoral digital dental impressions with iTero and extraoral digitization with the iTero and a model scanner. Am J Orthod Dentofacial Orthop. 2013; 144(3):471-8. DOI: 10.1016/j.ajodo.2013.04.017. View