Influence of Object Translucency on the Scanning Accuracy of a Powder-free Intraoral Scanner: A Laboratory Study

Overview
Journal J Prosthet Dent
Specialty Dentistry
Date 2016 Jul 28
PMID 27460324
Citations 24
Authors
Affiliations
Soon will be listed here.
Abstract

Statement Of Problem: Limited information is available regarding the influence of object translucency on the scanning accuracy of a powder-free intraoral scanner.

Purpose: The purpose of this in vitro study was to evaluate the scanning accuracy of a confocal microscopy principle powder-free intraoral scanner on ceramic copings and to analyze the relationship between scanning accuracy and object translucency.

Methods: Six slice specimens (12×10 mm) and 6 offset copings (1.00-mm thickness) were made from different translucent homogeneous ceramic blocks (CEREC Blocs, S0-M to S5-M, highest to lowest translucency). The primary sintered zirconia offset coping was produced in the same way as the control. Optical parameters related to the translucency of each slice were measured with a spectrophotometer. Three-dimensional (3D) datasets of the surface morphology of offset copings were obtained by using the intraoral scanner. The same white wax resin bases were used for registration. Quantitative parameters of scanning trueness and precision were measured. One-way ANOVA was used to analyze the values of each parameter among the 6 ceramic blocks. Bivariate correlation was used to analyze the relationships between each parameter of scanning accuracy and translucency (α=.05).

Results: Translucent copings showed a positive 3D bias (S0-M to S5-M: 0.149 ±0.038 mm to 0.068 ±0.020 mm), a narrower collar diameter (Dd=-0.067 mm), larger convergence angle (ΔΦ=2.79 degrees), and larger curvature radius of the internal gingivoaxial corner (Δρ=0.236 mm). The smaller the percentage sum of scattering and absorption, the greater was the occurrence of scanning bias (r=-0.918) and curvature (r=-0.935) decrease.

Conclusions: Use of the tested powder-free intraoral scanner, higher translucency objects (greater translucency than S1-M/A1C) resulted in lower scanning accuracy and morphological changes. Therefore, more suitable methods of measurement are still required.

Citing Articles

Comparison of intraoral and extraoral scanners for volumetric assessment before and after caries removal by the ICDAS score: a quantitative analysis.

Kanar O, Tagtekin D, Korkut B Clin Oral Investig. 2024; 28(11):624.

PMID: 39495394 DOI: 10.1007/s00784-024-06019-0.


Evaluating the accuracy of CEREC intraoral scanners for inlay restorations: impact of adjacent tooth materials.

Kwon Y, Kim J, Park J, Son S BMC Oral Health. 2024; 24(1):1033.

PMID: 39227885 PMC: 11370093. DOI: 10.1186/s12903-024-04794-6.


Aspects of Occlusal Recordings Performed with the T-Scan System and with the Medit Intraoral Scanner.

Popa A, Vladutu D, Turcu A, Tartea D, Ionescu M, Paunescu C Diagnostics (Basel). 2024; 14(13).

PMID: 39001349 PMC: 11241177. DOI: 10.3390/diagnostics14131457.


Evaluation of the effect of different core substrates on the accuracy of intraoral scanners.

Khoshkhahesh M, Enteghad S, Aghasadeghi K, Farzin M, Taghva M, Mosadad S Clin Exp Dent Res. 2024; 10(3):e899.

PMID: 38752461 PMC: 11097246. DOI: 10.1002/cre2.899.


Expediting the Rehabilitation of Severely Resorbed Ridges Using a Combination of CAD-CAM and Analog Techniques: A Case Report.

Jurado C, Azpiazu-Flores F, Fu C, Rojas-Rueda S, Guzman-Perez G, Floriani F Medicina (Kaunas). 2024; 60(2).

PMID: 38399547 PMC: 10890117. DOI: 10.3390/medicina60020260.