Abstract

The aim of this study was to assess the accuracy of an algorithm for automated measurement of left ventricular ejection fraction (LVEF) available on handheld ultrasound devices (HUDs). One hundred twelve patients admitted to the cardiology department underwent assessment performed with an HUD. In each case, the four-chamber apical view was obtained, and LVEF was calculated with LVivo software. Subsequently, during the examination performed with the use of the stationary echocardiograph, the 3-D measurement of LVEF was recorded. The average LVEFs measured with LVivo and the 3-D reference method were 46 ± 14% and 48 ± 14%, respectively. The correlation between the measurements obtained with the HUD and 3-D evaluation was high (r = 0.92, 95% confidence interval: 0.87-0.95, p < 0.0001). The mean difference between the LVEF obtained with LVivo and the 3-D LVEF was not significant (mean difference: -0.61%, 95% confidence interval: -1.89 to 0.68, p = 0.31). The LVivo software despite its limitations is capable of the accurate LVEF measurement when the acquired views are of at least good imaging quality.

Citing Articles

Reliability of spectral Doppler in handheld ultrasonographic device.

Filipiak-Strzecka D, Kasprzak J, Lipiec P Int J Cardiovasc Imaging. 2025; .

PMID: 40063157 DOI: 10.1007/s10554-025-03372-3.


Evaluation of left ventricular ejection fraction by a new automatic tool on a pocket ultrasound device: Concordance study with cardiac magnetic resonance imaging.

Berger L, Coisy F, Sammoud S, de Oliveira F, Grandpierre R, Grau-Mercier L PLoS One. 2024; 19(8):e0308580.

PMID: 39133705 PMC: 11318925. DOI: 10.1371/journal.pone.0308580.


The importance of patient characteristics, operators, and image quality for the accuracy of heart failure diagnosis by general practitioners using handheld ultrasound devices.

Magelssen M, Hjorth-Hansen A, Andersen G, Graven T, Kleinau J, Skjetne K Eur Heart J Imaging Methods Pract. 2024; 2(1):qyad047.

PMID: 39045176 PMC: 11195733. DOI: 10.1093/ehjimp/qyad047.


Artificial intelligence-assisted evaluation of cardiac function by oncology staff in chemotherapy patients.

Papadopoulou S, Dionysopoulos D, Mentesidou V, Loga K, Michalopoulou S, Koukoutzeli C Eur Heart J Digit Health. 2024; 5(3):278-287.

PMID: 38774364 PMC: 11104473. DOI: 10.1093/ehjdh/ztae017.


Artificial Intelligence-Powered Left Ventricular Ejection Fraction Analysis Using the LVivoEF Tool for COVID-19 Patients.

Dadon Z, Steinmetz Y, Levi N, Orlev A, Belman D, Butnaru A J Clin Med. 2023; 12(24).