3D-printed, Patient-specific Intracranial Aneurysm Models: From Clinical Data to Flow Experiments with Endovascular Devices

Overview
Journal Med Phys
Specialty Biophysics
Date 2021 Jan 11
PMID 33428778
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Flow models of intracranial aneurysms (IAs) can be used to test new and existing endovascular treatments with flow modulation devices (FMDs). Additionally, 4D flow magnetic resonance imaging (MRI) offers the ability to measure hemodynamics. This way, the effect of FMDs can be determined noninvasively and compared to patient data. Here, we describe a cost-effective method for producing flow models to test the efficiency of FMDs with 4D flow MRI.

Methods: The models were based on human radiological data (internal carotid and basilar arteries) and printed in 3D with stereolithography. The models were printed with three different printing layers (25, 50, and 100 µm thickness). To evaluate the models in vitro, 3D rotational angiography, time-of-flight MRI, and 4D flow MRI were employed. The flow and geometry of one model were compared with in vivo data. Two FMDs (FMD1 and FMD2) were deployed into two different IA models, and the effect on the flow was estimated by 4D flow MRI.

Results: Models printed with different layer thicknesses exhibited similar flow and little geometric variation. The mean spatial difference between the vessel geometry measured in vivo and in vitro was 0.7 ± 1.1 mm. The main flow features, such as vortices in the IAs, were reproduced. The velocities in the aneurysms were similar in vivo and in vitro (mean velocity magnitude: 5.4 ± 7.6 and 7.7 ± 8.6 cm/s, maximum velocity magnitude: 72.5 and 55.1 cm/s). By deploying FMDs, the mean velocity was reduced in the IAs (from 8.3 ± 10 to 4.3 ± 9.32 cm/s for FMD1 and 9.9 ± 12.1 to 2.1 ± 5.6 cm/s for FMD2).

Conclusions: The presented method allows to produce neurovascular models in approx. 15 to 30 h. The resulting models were found to be geometrically accurate, reproducing the main flow patterns, and suitable for implanting FMDs as well as 4D flow MRI.

Citing Articles

A Method for Developing a Free-Standing Cerebrovascular Lumen Model.

Kurokawa T, Tanaka Y, Umemura T, Yamamoto J J Neuroendovasc Ther. 2025; 19(1).

PMID: 40018277 PMC: 11865168. DOI: 10.5797/jnet.tn.2024-0085.


Interleaved flow-sensitive dephasing (iFSD): Toward enhanced blood flow suppression and preserved T weighting and overall signals in 3D TSE-based neuroimaging.

Kong Q, Xiao J, Shiroishi M, Sheikh-Bahaei N, Cen S, Khatibi K Magn Reson Med. 2024; 93(5):1911-1923.

PMID: 39648519 PMC: 11893033. DOI: 10.1002/mrm.30391.


Wallenberg Syndrome Secondary to Vertebrobasilar Aneurysm Associated With Subclavian Steal Syndrome.

Castaneda Aguayo F, Aguirre A, Garcia I, Reyes Soto G, Catillo-Rangel C, Castillo Soriano C Cureus. 2024; 16(10):e72623.

PMID: 39610625 PMC: 11604026. DOI: 10.7759/cureus.72623.


Design, manufacturing, and multi-modal imaging of stereolithography 3D printed flexible intracranial aneurysm phantoms.

Yalman A, Jafari A, Leger E, Mastroianni M, Teimouri K, Savoji H Med Phys. 2024; 52(2):742-749.

PMID: 39546636 PMC: 11788251. DOI: 10.1002/mp.17518.


Non-invasive follow-up for intracranial aneurysms treated with contour neurovascular system-comparison of digital subtraction angiography (DSA) to magnetic resonance imaging (MRI) and spectral computed tomography angiography (CTA) in vitro.

Madjidyar J, Pravdivtseva M, Hensler J, Jansen O, Larsen N, Wodarg F Interv Neuroradiol. 2024; :15910199241277907.

PMID: 39219551 PMC: 11569776. DOI: 10.1177/15910199241277907.