» Articles » PMID: 33384657

Possibility of Worsening Flow Diversion Effect Due to Morphological Changes of a Stented Artery With Multiple Overlapping Stents for Partially Thrombosed Vertebral Artery Aneurysms

Overview
Journal Front Neurol
Specialty Neurology
Date 2021 Jan 1
PMID 33384657
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Morphological changes of a stented artery can cause a flow diversion effect to reduce intra-aneurysmal flow; however, there is a potential for the negative effect of increased intra-aneurysmal flow. We present cases with multiple overlapping stents for a partially thrombosed vertebral artery aneurysm and characterize the hemodynamic properties of a recurrent case by focusing on the morphological changes of the stented artery. Between October 2017 and April 2019, four consecutive cases of symptomatic unruptured large and giant partially thrombosed vertebral artery aneurysms were treated with multiple overlapping low-profile visualized intraluminal support stents and no coils. Both angiographic and clinical outcomes were assessed. Computational fluid dynamics analysis was performed to clarify hemodynamic features. The degree of pressure elevation was calculated as the pressure difference (Pd). Wall shear stress (WSS) was also calculated. In three of the four cases, successful flow reduction was achieved with no morphological change of the stented arteries. The patients' symptoms were gradually improved. The remaining case required additional stents after the initial treatment. In the recurrent case, Pd was noticeably elevated at the aneurysm neck after treatment, and WSS was generally increased in the area due to altered blood flow into the aneurysm dome caused by morphological changes of the stented artery. Overlapping stents can be used for the treatment of large and giant thrombosed vertebral artery aneurysms with flow diversion effect; however, morphological changes of the stented artery requires careful attention as it may lead to an increase in the intra-aneurysmal flow, causing negative outcomes.

Citing Articles

A Giant Stent for Giant Cerebral Aneurysms-The Accero-Rex-Stent.

Kraehling H, Akkurt B, Elsharkawy M, Ayad A, Ergawy M, Celik E J Clin Med. 2024; 13(2).

PMID: 38256521 PMC: 10816579. DOI: 10.3390/jcm13020388.


Clinical and Angiographic Outcomes of Endovascular Treatment for Acute Intracranial Vertebral Artery Dissecting Aneurysms Using Double-Overlapping Stents : Low-Profile Visualized Intraluminal Support within Enterprise Stents.

Koo J, Hwang E, Song J, Lim Y J Korean Neurosurg Soc. 2023; 67(5):531-540.

PMID: 38130141 PMC: 11375064. DOI: 10.3340/jkns.2023.0224.


Flow Redirection Endoluminal Device Flow Diverter Placement for Recurrent Dolichoectatic Vertebrobasilar Artery Aneurysm Treated with Multiple Low-Profile Visualized Intraluminal Support Stents: A Case Study.

Hosogai M, Okazaki T, Sakamoto S, Ishii D, Kuwabara M, Horie N J Neuroendovasc Ther. 2023; 17(1):27-31.

PMID: 37501885 PMC: 10370513. DOI: 10.5797/jnet.cr.2022-0045.


Progressive volume reduction and long-term aneurysmal collapse following flow diversion treatment of giant and symptomatic cerebral aneurysms.

Sirakova K, Penkov M, Matanov S, Minkin K, Ninov K, Hadzhiyanev A Front Neurol. 2022; 13:972599.

PMID: 36034286 PMC: 9403733. DOI: 10.3389/fneur.2022.972599.


Hemodynamic Comparison of Treatment Strategies for Intracranial Vertebral Artery Fusiform Aneurysms.

Jiang Y, Lu G, Ge L, Zou R, Li G, Wan H Front Neurol. 2022; 13:927135.

PMID: 35873788 PMC: 9296783. DOI: 10.3389/fneur.2022.927135.


References
1.
Luo B, Yang X, Wang S, Li H, Chen J, Yu H . High shear stress and flow velocity in partially occluded aneurysms prone to recanalization. Stroke. 2011; 42(3):745-53. DOI: 10.1161/STROKEAHA.110.593517. View

2.
Wang C, Tian Z, Liu J, Jing L, Paliwal N, Wang S . Flow diverter effect of LVIS stent on cerebral aneurysm hemodynamics: a comparison with Enterprise stents and the Pipeline device. J Transl Med. 2016; 14(1):199. PMC: 4930570. DOI: 10.1186/s12967-016-0959-9. View

3.
Yan Z, Zheng K, Xiong Y, Lan F, Wang Y, Tan X . Intracranial Complex Ruptured Aneurysms Coiled with Overlapping Low-Profile Visualized Intraluminal Support Stents: Another Available Option for Complex Ruptured Intracranial Aneurysms. World Neurosurg. 2019; 125:e22-e28. DOI: 10.1016/j.wneu.2018.12.142. View

4.
Ishii A, Chihara H, Kikuchi T, Arai D, Ikeda H, Miyamoto S . Contribution of the straightening effect of the parent artery to decreased recanalization in stent-assisted coiling of large aneurysms. J Neurosurg. 2016; 127(5):1063-1069. DOI: 10.3171/2016.9.JNS16501. View

5.
Cebral J, Mut F, Raschi M, Scrivano E, Ceratto R, Lylyk P . Aneurysm rupture following treatment with flow-diverting stents: computational hemodynamics analysis of treatment. AJNR Am J Neuroradiol. 2010; 32(1):27-33. PMC: 7964947. DOI: 10.3174/ajnr.A2398. View