» Articles » PMID: 33204849

Continuous Noninvasive Monitoring of Arterial Pressure Using the Vascular Unloading Technique in Comparison to the Invasive Gold Standard in Elderly Comorbid Patients: A Prospective Observational Study

Overview
Journal Health Sci Rep
Specialty General Medicine
Date 2020 Nov 18
PMID 33204849
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Aims: Elderly patients aged ≥65 years represent a growing population in the perioperative field, particularly orthopedic and vascular surgery. The higher degree of age-related or comorbid-dependent vascular alterations renders these patients at risk for hemodynamic complications and likely denote a possible limitation for modern, non-invasive arterial pressure monitoring devices. The aim was to compare vascular unloading technique-derived to invasive measurements of systolic (SAP), diastolic (DAP), and mean arterial pressure (MAP) in elderly perioperative patients.

Methods: This prospective observational study included patients aged ≥65 years scheduled for orthopedic and patients ≥50 years with peripheral artery disease Fontaine stage ≥ II scheduled for vascular surgery, respectively. Invasive radial artery and non-invasive finger-cuff (Nexfin system) arterial pressures were recorded before and after induction of general anesthesia and during surgery. Correlation, Bland-Altman, and concordance analyses were performed. Measurements of arterial pressure were also compared during intraoperative hypotension (MAP <70 mm Hg) and hypertension (MAP >105 mm Hg).

Results: Sixty patients with orthopedic (N = 25, mean (SD) age 77 (5) years) and vascular surgery (N = 35, age 69 [10] years) were enrolled. Seven hundred data pairs of all patients were analysed and pooled bias and percentage error were: SAP: 14.43 mm Hg, 43.79%; DAP: -2.40 mm Hg, 53.78% and MAP: 1.73 mm Hg, 45.05%. Concordance rates were 84.01% for SAP, 77.87% for DAP, and 86.47% for MAP. Predefined criteria for interchangeability of absolute and trending values could neither be reached in the overall nor in the subgroup analyses orthopedic vs vascular surgery. During hypertension, percentage error was found to be lowest for all pressure values, still not reaching predefined criteria.

Conclusion: Arterial pressure monitoring with the vascular unloading technique did not reach criteria of interchangeability for absolute and trending values. Nevertheless, the putatively beneficial use of noninvasive arterial pressure measurements should be further evaluated in the elderly perioperative patient.

Citing Articles

Comparison of the ClearSight™ finger cuff monitor versus invasive arterial blood pressure measurement in elective cardiac surgery patients: a prospective observational study.

Goncin U, Liu K, Rawlyk B, Dalkilic S, Walker M, Norton J Can J Anaesth. 2024; 71(11):1495-1504.

PMID: 39317830 DOI: 10.1007/s12630-024-02834-x.


Comparison of oscillometric, non-invasive and invasive arterial pressure monitoring in patients undergoing laparoscopic bariatric surgery - a secondary analysis of a prospective observational study.

Hansen J, Pohlmann M, Beckmann J, Klose P, Gruenewald M, Renner J BMC Anesthesiol. 2022; 22(1):83.

PMID: 35346046 PMC: 8962134. DOI: 10.1186/s12871-022-01619-3.


Continuous noninvasive monitoring of arterial pressure using the vascular unloading technique in comparison to the invasive gold standard in elderly comorbid patients: A prospective observational study.

Klose P, Lorenzen U, Berndt R, Borzikowsky C, Hill M, Gruenewald M Health Sci Rep. 2020; 3(4):e204.

PMID: 33204849 PMC: 7654630. DOI: 10.1002/hsr2.204.

References
1.
Alfano G, Fontana F, Cappelli G . Noninvasive Blood Pressure Measurement in Maintenance Hemodialysis Patients: Comparison of Agreement between Oscillometric and Finger-Cuff Methods. Nephron. 2017; 136(4):309-317. DOI: 10.1159/000457812. View

2.
Rogge D, Nicklas J, Schon G, Grothe O, Haas S, Reuter D . Continuous Noninvasive Arterial Pressure Monitoring in Obese Patients During Bariatric Surgery: An Evaluation of the Vascular Unloading Technique (Clearsight system). Anesth Analg. 2019; 128(3):477-483. DOI: 10.1213/ANE.0000000000003943. View

3.
Critchley L, Critchley J . A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 2003; 15(2):85-91. DOI: 10.1023/a:1009982611386. View

4.
Ramsey 3rd M . Noninvasive automatic determination of mean arterial pressure. Med Biol Eng Comput. 1979; 17(1):11-8. DOI: 10.1007/BF02440948. View

5.
Biais M, Stecken L, Ottolenghi L, Roullet S, Quinart A, Masson F . The ability of pulse pressure variations obtained with CNAP™ device to predict fluid responsiveness in the operating room. Anesth Analg. 2011; 113(3):523-8. DOI: 10.1213/ANE.0b013e3182240054. View