» Articles » PMID: 33134222

Noise Reduction in CT Images Using a Selective Mean Filter

Overview
Date 2020 Nov 2
PMID 33134222
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Noise reduction is a method for reducing CT dose; however, it can reduce image quality.

Objective: This study aims to propose a selective mean filter (SMF) and evaluate its effectiveness for noise suppression in CT images.

Material And Methods: This experimental study proposed and implemented the new noise reduction algorithm. The proposed algorithm is based on a mean filter (MF), but the calculation of the mean pixel value using the neighboring pixels in a kernel selectively applied a threshold value based on the noise of the image. The SMF method was evaluated using images of phantoms. The dose reduction was estimated by comparing the image noise acquired with a lower dose after implementing the SMF method and the noise in the original image acquired with a higher dose. For comparison, the images were also filtered with an adaptive mean filter (AMF) and a bilateral filter (BF).

Results: The spatial resolution of the image filtered with the SMF was similar to the original images and the images filtered with the BF. While using the AMF, spatial resolution was significantly corrupted. The noise reduction achieved using the SMF was up to 75%, while it was up to 50% using the BF.

Conclusion: SMF significantly reduces the noise and preserves the spatial resolution of the image. The noise reduction was more pronounced with BF, and less pronounced with AMF.

Citing Articles

LightGBM is an Effective Predictive Model for Postoperative Complications in Gastric Cancer: A Study Integrating Radiomics with Ensemble Learning.

Wang W, Sheng R, Liao S, Wu Z, Wang L, Liu C J Imaging Inform Med. 2024; 37(6):3034-3048.

PMID: 38940888 PMC: 11612084. DOI: 10.1007/s10278-024-01172-0.


Ensemble classification of integrated CT scan datasets in detecting COVID-19 using feature fusion from contourlet transform and CNN.

Nur-A-Alam M, Nasir M, Ahsan M, Based M, Haider J, Kowalski M Sci Rep. 2023; 13(1):20063.

PMID: 37973820 PMC: 10654719. DOI: 10.1038/s41598-023-47183-9.


Automated development of the contrast-detail curve based on statistical low-contrast detectability in CT images.

Anam C, Naufal A, Fujibuchi T, Matsubara K, Dougherty G J Appl Clin Med Phys. 2022; 23(9):e13719.

PMID: 35808971 PMC: 9512356. DOI: 10.1002/acm2.13719.


Elimination of biological and physical artifacts in abdomen and brain computed tomography procedures using filtering techniques.

Omer H, Tamam N, Alameen S, Algadi S, Tai D, Sulieman A Saudi J Biol Sci. 2022; 29(4):2180-2186.

PMID: 35531247 PMC: 9073048. DOI: 10.1016/j.sjbs.2021.11.043.


Effect Evaluation of Platelet-Rich Plasma Combined with Vacuum Sealing Drainage on Serum Inflammatory Factors in Patients with Pressure Ulcer by Intelligent Algorithm-Based CT Image.

Yang J, Xiao C, Wen H, Sun K, Wu X, Feng X Comput Math Methods Med. 2022; 2022:8916076.

PMID: 35281950 PMC: 8906978. DOI: 10.1155/2022/8916076.

References
1.
Mazrani W, McHugh K, Marsden P . The radiation burden of radiological investigations. Arch Dis Child. 2007; 92(12):1127-31. PMC: 2066089. DOI: 10.1136/adc.2006.101782. View

2.
Christianson O, Winslow J, Frush D, Samei E . Automated Technique to Measure Noise in Clinical CT Examinations. AJR Am J Roentgenol. 2015; 205(1):W93-9. DOI: 10.2214/AJR.14.13613. View

3.
Kalra M, Maher M, Sahani D, Blake M, Hahn P, Avinash G . Low-dose CT of the abdomen: evaluation of image improvement with use of noise reduction filters pilot study. Radiology. 2003; 228(1):251-6. DOI: 10.1148/radiol.2281020693. View

4.
Anam C, Arif I, Haryanto F, Lestari F, Widita R, Budi W . An Improved Method of Automated Noise Measurement System in CT Images. J Biomed Phys Eng. 2021; 11(2):163-174. PMC: 8064134. DOI: 10.31661/jbpe.v0i0.1198. View

5.
Greess H, Lutze J, Nomayr A, Wolf H, Hothorn T, Kalender W . Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation. Eur Radiol. 2004; 14(6):995-9. DOI: 10.1007/s00330-004-2301-9. View