Anterior Chamber Angle Imaging with Swept-source Optical Coherence Tomography: Comparison Between CASIAII and ANTERION

Overview
Journal Sci Rep
Specialty Science
Date 2020 Oct 31
PMID 33127944
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

This study compared the test-retest variabilities and measurement agreement of anterior chamber angle (ACA) dimensions measured by two anterior segment swept-source optical coherence tomography (SS-OCT)-the ANTERION (Heidelberg Engineering, Heidelberg, Germany) and CASIAII (Tomey, Nagoya, Japan). Thirty-eight subjects, 18 patients with primary angle closure and 20 healthy participants with open angles, were included. The mean age was 54.7 ± 15.8 years (range: 26-75 years). One eye of each subject was randomly selected for anterior segment imaging by ANTERION and CASIAII, using the same scan pattern (6 evenly spaced radial scans across the anterior segment for three times) in the same visit. The between- and within-instrument agreement and repeatability coefficients of angle open distance (AOD500), trabecular-iris space area (TISA500), lens vault (LV), scleral spur-scleral spur distance (SSD), anterior chamber depth (ACD), and pupil diameter (PD) were measured. The anterior and posterior boundaries of the cornea, iris, and lens were automatically segmented by the SS-OCT instruments; the scleral spur was manually located by a single masked observer. There were significant differences between ANTERION and CASIAII measurements; the SSD, PD, and ACD were smaller whereas AOD500 and TISA500 were greater in ANTERION compared with CASIAII (P < 0.001). Anterior segment measurements obtained from the two SS-OCT instruments showed strong associations (R ranged between 0.866 and 0.998) although the between-instrument agreement was poor; the spans of 95% limits of between-instrument agreement were ≥ 1.5-folds than the within-instrument agreement for either instrument. Whereas both SS-OCT instruments showed low test-retest measurement variabilities, the repeatability coefficients of AOD500, TISA500, ACD, and PD were slightly smaller for CASIAII than ANTERION (P ≤ 0.012).

Citing Articles

Loss values of style transfer from UBM to AS-OCT images for plateau iris classification.

Kaothanthong N, Wanichwecharungruang B, Chantangphol P, Pattanapongpaiboon W, Theeramunkong T Sci Rep. 2024; 14(1):31157.

PMID: 39732818 PMC: 11682431. DOI: 10.1038/s41598-024-82327-5.


Swept-source optical coherence tomography in ocular surface diseases: anterior segment analysis repeatability and its limits.

Recchioni A, Venkataraman A, Rauz S, Dominguez-Vicent A Front Med (Lausanne). 2024; 11:1385294.

PMID: 39156694 PMC: 11327131. DOI: 10.3389/fmed.2024.1385294.


Impact of lens autofluorescence and opacification on retinal imaging.

von der Emde L, Rennen G, Vaisband M, Hasenauer J, Liegl R, Fleckenstein M BMJ Open Ophthalmol. 2024; 9(1).

PMID: 38684375 PMC: 11086461. DOI: 10.1136/bmjophth-2023-001628.


Agreement and Reproducibility of Anterior Chamber Angle Measurements between CASIA2 Built-In Software and Human Graders.

Espinoza G, Iglesias K, Parra J, Rodriguez-Una I, Serrano-Gomez S, Prada A J Clin Med. 2023; 12(19).

PMID: 37835024 PMC: 10573880. DOI: 10.3390/jcm12196381.


Iridociliary measurements using the Anterion swept-source optical coherence tomography.

Patyal S, Yadav A, Anthony H, Ansari F Indian J Ophthalmol. 2023; 71(1):109-112.

PMID: 36588218 PMC: 10155553. DOI: 10.4103/ijo.IJO_1779_22.


References
1.
Radhakrishnan S, Huang D, Smith S . Optical coherence tomography imaging of the anterior chamber angle. Ophthalmol Clin North Am. 2005; 18(3):375-81, vi. DOI: 10.1016/j.ohc.2005.05.007. View

2.
Nongpiur M, He M, Amerasinghe N, Friedman D, Tay W, Baskaran M . Lens vault, thickness, and position in Chinese subjects with angle closure. Ophthalmology. 2010; 118(3):474-9. DOI: 10.1016/j.ophtha.2010.07.025. View

3.
Ang M, Baskaran M, Werkmeister R, Chua J, Schmidl D, Aranha Dos Santos V . Anterior segment optical coherence tomography. Prog Retin Eye Res. 2018; 66:132-156. DOI: 10.1016/j.preteyeres.2018.04.002. View

4.
Leung C, Weinreb R . Anterior chamber angle imaging with optical coherence tomography. Eye (Lond). 2011; 25(3):261-7. PMC: 3178313. DOI: 10.1038/eye.2010.201. View

5.
Thomas R . Anterior segment optical coherence tomography. Ophthalmology. 2007; 114(12):2362-3. DOI: 10.1016/j.ophtha.2007.05.050. View