Efficacy and Safety of High-dose Ultrasound Cyclo-plasty Procedure in Refractory Glaucoma

Overview
Specialty Ophthalmology
Date 2020 Sep 21
PMID 32953577
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: To evaluate the efficacy and safety of high-dose ultrasound cyclo-plasty (UCP) for the treatment of refractory glaucoma in Chinese patients.

Methods: In this 6-month retrospective study, 37 eyes of 37 patients suffering from severe glaucoma with uncontrolled intraocular pressure (IOP) of ≥21 mm Hg underwent 8-s ultrasonic cyclocoagulation with ten active piezoelectric elements. A complete ophthalmic examination was performed before and at 1d, 1, 3, 6mo after UCP. Therapeutic success was defined as IOP reduction from baseline ≥20% and IOP ≥5 mm Hg without adding new glaucoma medication compare to baseline at the 6-month follow-up visit. In addition to mean IOP at each follow-up visit, medications used and complications were also detected and compared to baseline.

Results: After UCP procedure, the mean IOP was significantly reduced (<0.01) from the preoperative 44.1±11.9 mm Hg to postoperative 26.7±11.8 mm Hg at 3mo, and 30.4±14.5 mm Hg at 6mo. The overall mean IOP reductions achieved at 3 and 6mo were 39% and 31% compared to baseline IOP. Sixty-one percent of patients responded well to UCP treatment with a mean IOP reduction of 48% at 3mo and 42% at 6mo. Ocular pain in most of patients were alleviated. No serious intraoperative or postoperative complications occurred.

Conclusion: High-dose UCP treatment is an effective and safe procedure to reduce IOP in Chinese patients with severe glaucoma.

Citing Articles

Iris neovascularization and neurotrophic keratopathy following ultrasound cycloplasty in refractory glaucoma: case series.

Chen Z, Wu Y, Yu M BMC Ophthalmol. 2024; 24(1):188.

PMID: 38654174 PMC: 11040811. DOI: 10.1186/s12886-024-03460-3.


Efficacy and safety of ultrasound cycloplasty for the treatment of glaucoma: a Meta-analysis.

Wu T, Yin X, Li J, Lu P Int J Ophthalmol. 2023; 16(8):1317-1325.

PMID: 37602344 PMC: 10398525. DOI: 10.18240/ijo.2023.08.19.


Efficacy and safety of high-intensity focused ultrasound cyclo-plasty in glaucoma.

Chen D, Guo X, Luo S, Lu Y, Tang X BMC Ophthalmol. 2022; 22(1):401.

PMID: 36207670 PMC: 9542456. DOI: 10.1186/s12886-022-02622-5.


Ultrasound Cyclo Plasty for Treatment of Surgery-Naïve Open-Angle Glaucoma Patients: A Prospective, Multicenter, 2-Year Follow-Up Trial.

Figus M, Posarelli C, Nardi M, Stalmans I, Vandewalle E, Melamed S J Clin Med. 2021; 10(21).

PMID: 34768500 PMC: 8584324. DOI: 10.3390/jcm10214982.

References
1.
Yamashita H, Sears M . Proof that the ciliary epithelium can regenerate. Exp Eye Res. 1978; 27(2):199-213. DOI: 10.1016/0014-4835(78)90089-1. View

2.
Aptel F, Tadjine M, Rouland J . Efficacy and Safety of Repeated Ultrasound Cycloplasty Procedures in Patients With Early or Delayed Failure After a First Procedure. J Glaucoma. 2019; 29(1):24-30. DOI: 10.1097/IJG.0000000000001400. View

3.
Aptel F, Charrel T, Lafon C, Romano F, Chapelon J, Blumen-Ohana E . Miniaturized high-intensity focused ultrasound device in patients with glaucoma: a clinical pilot study. Invest Ophthalmol Vis Sci. 2011; 52(12):8747-53. DOI: 10.1167/iovs.11-8137. View

4.
Mastropasqua R, Fasanella V, Mastropasqua A, Ciancaglini M, Agnifili L . High-Intensity Focused Ultrasound Circular Cyclocoagulation in Glaucoma: A Step Forward for Cyclodestruction?. J Ophthalmol. 2017; 2017:7136275. PMC: 5420440. DOI: 10.1155/2017/7136275. View

5.
Isard P, Mentek M, Clement D, Begle A, Romano F, Aptel F . High intensity focused ultrasound cyclocoagulation in dogs with primary glaucoma: a preliminary study. Open Vet J. 2018; 8(3):305-312. PMC: 6102420. DOI: 10.4314/ovj.v8i3.9. View