» Articles » PMID: 32865039

Short Single-wedge Stems Have Higher Risk of Periprosthetic Fracture Than Other Cementless Stem Designs in Dorr Type A Femurs: a Finite Element Analysis

Overview
Journal Hip Int
Specialty Orthopedics
Date 2020 Sep 1
PMID 32865039
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The use of total hip arthroplasty (THA) femoral stems that transmit force in a favourable manner to the femur may minimise periprosthetic fractures. Finite element analysis (FEA) is a computerised method that analyses the effect of forces applied to a structure with complex shape. Our aim was to apply FEA to compare primary THA cementless stem designs and their association with periprosthetic fracture risk.

Methods: 3-dimensional (3D) models of a Dorr Type A femur and 5 commonly used primary THA cementless stem designs (short single wedge, standard-length single wedge, modular, double-wedge metaphyseal filling, and cylindrical fully coated) were developed using computed tomography (CT) imaging. Implant insertion, single-leg stance, and twisting with a planted foot were simulated. FEA was performed, and maximum femoral strain along the implant-bone interface recorded.

Results: Femoral strain was highest with short single-wedge stem design (0.3850) and lowest with standard-length single-wedge design (0.0520). Location of maximum femoral strain varied by stem design, but not with implant insertion, single-leg stance, or twisting with a planted foot. Strain was as high during implant insertion as with single-leg stance or twisting with a planted foot.

Conclusions: Our results suggest the risk of intraoperative and postoperative periprosthetic fracture with THA in a Dorr A femur is highest with short single-wedge stems and lowest with standard-length single-wedge stems. Consideration may be given to minimising the use of short single-wedge stems in THA. Implant-specific sites of highest strain should be carefully inspected for fracture.

Citing Articles

CORR Insights®: Are There Differences in Performance Among Femoral Stem Brands Utilized in Cementless Hemiarthroplasty for Treatment of Geriatric Femoral Neck Fractures?.

Keeney J Clin Orthop Relat Res. 2024; 483(2):265-267.

PMID: 39330916 PMC: 11753741. DOI: 10.1097/CORR.0000000000003266.


Callus formation after total hip arthroplasty using a short tapered-wedge stem.

Ishii S, Baba T, Shirogane Y, Hayashi K, Homma Y, Muto O Skeletal Radiol. 2024; 54(3):447-455.

PMID: 39052073 DOI: 10.1007/s00256-024-04756-7.


Stress reduction through cortical bone thickening improves bone mechanical behavior in adult female Beclin-1 mice.

Yang J, Pei Q, Wu X, Dai X, Li X, Pan J Front Bioeng Biotechnol. 2024; 12:1357686.

PMID: 38600946 PMC: 11004267. DOI: 10.3389/fbioe.2024.1357686.


Biomechanical study of different bone cement distribution on osteoporotic vertebral compression Fracture-A finite element analysis.

Zhou C, Meng X, Huang S, Chen H, Zhou H, Liao Y Heliyon. 2024; 10(5):e26726.

PMID: 38434291 PMC: 10907677. DOI: 10.1016/j.heliyon.2024.e26726.


Multi-Scale Topology Optimization of Femoral Stem Structure Subject to Stress Shielding Reduce.

Xiao Z, Wu L, Wu W, Tang R, Dai J, Zhu D Materials (Basel). 2023; 16(8).