Receptor Antibody Time-resolved A2 Phospholipase Bead Immunochromatography and Its Application in Idiopathic Membranous Nephropathy

Overview
Journal J Clin Lab Anal
Publisher Wiley
Date 2020 Aug 9
PMID 32767448
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To detect phospholipase A2 receptor (PLA2R) antibody by established time-resolved fluorescent bead immunochromatographic assay.

Methods: The reaction time of coupling, pH of the reaction, and coupling ratio of the label to PLA2R were determined. The EDC method was used to covalently couple PLA2R to time-resolved fluorescent beads, which were sprayed onto a bonding pad. PLA2R and rabbit anti-PLA2R antibody sprayed onto a nitrocellulose membrane were used as detection and quality control lines, respectively. Immunochromatographic test strips were prepared to enable rapid detection of PLA2R antibodies. Various technical indicators were evaluated, and the correlation among this method, enzyme-linked immunosorbent assay (ELISA), and serum analysis was examined.

Results: The pH suitable for labeling was 6.5. The optimal mass ratio of PLA2R protein to fluorescent beads was 0.08:1, and the reaction time of coupling was at least 1.5 hours. The appropriate spray film size of the coupled fluorescent bead was 5 μL/cm, and the appropriate staining concentration of the test line was 0.28 mg/mL. Further, 80 µL of sample was required for the test, and the result was obtained in only 15 minutes. The measurable range of this method was 5-1500 RU/mL. Intra- and inter-assay coefficients of variation were 7.61% and 11.07%, respectively, with an average recovery rate of 93.77%. The method showed a good correlation with ELISA, with a correlation coefficient of 0.936.

Conclusions: This method could better meet the clinical demand for idiopathic membranous nephropathy (IMN) detection.

Citing Articles

Establishment of trypsinogen-2 Amplification Luminescent Proximity Homogeneous Assay and its Application in Acute Pancreatitis.

Chen M, Fang H, Wu J, Huang Y, Cheng F, Qin Y J Fluoresc. 2024; .

PMID: 39235654 DOI: 10.1007/s10895-024-03917-0.


Establishment of time-resolved fluoroimmunoassay of IgG4 based on magnetic microspheres.

Wu Q, Hang C, Li T, Wang B, Qin Y, Wang Y J Clin Lab Anal. 2021; 35(8):e23874.

PMID: 34125973 PMC: 8373317. DOI: 10.1002/jcla.23874.


Receptor antibody time-resolved A2 phospholipase bead immunochromatography and its application in idiopathic membranous nephropathy.

Huang B, Yang X, Zhang W, Wu J, Liu P, Hu Z J Clin Lab Anal. 2020; 34(12):e23508.

PMID: 32767448 PMC: 7755819. DOI: 10.1002/jcla.23508.

References
1.
Beck Jr L, Bonegio R, Lambeau G, Beck D, Powell D, Cummins T . M-type phospholipase A2 receptor as target antigen in idiopathic membranous nephropathy. N Engl J Med. 2009; 361(1):11-21. PMC: 2762083. DOI: 10.1056/NEJMoa0810457. View

2.
Huang B, Zhang Y, Wang L, Xu W, Zhang J, Zhang Q . Phospholipase A2 Receptor Antibody IgG4 Subclass Improves Sensitivity and Specificity in the Diagnosis of Idiopathic Membranous Nephropathy. Kidney Blood Press Res. 2019; 44(4):848-857. DOI: 10.1159/000500456. View

3.
Qin W, Beck Jr L, Zeng C, Chen Z, Li S, Zuo K . Anti-phospholipase A2 receptor antibody in membranous nephropathy. J Am Soc Nephrol. 2011; 22(6):1137-43. PMC: 3103733. DOI: 10.1681/ASN.2010090967. View

4.
Huang B, Wang L, Zhang Y, Zhang J, Zhang Q, Xiao H . A novel Time-resolved Fluoroimmunoassay for the quantitative detection of Antibodies against the Phospholipase A2 Receptor. Sci Rep. 2017; 7:46096. PMC: 5387400. DOI: 10.1038/srep46096. View

5.
Bobart S, De Vriese A, Pawar A, Zand L, Sethi S, Giesen C . Noninvasive diagnosis of primary membranous nephropathy using phospholipase A2 receptor antibodies. Kidney Int. 2019; 95(2):429-438. DOI: 10.1016/j.kint.2018.10.021. View