» Articles » PMID: 32673618

Development of a Novel Chemiluminescence Immunoassay for the Detection of Procalcitonin

Overview
Publisher Elsevier
Date 2020 Jul 17
PMID 32673618
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To evaluate the analytical performance of our previously developed chemiluminescence immunoassay (CLIA) kit for the detection of procalcitonin (PCT) and compare with the results obtained using the Vidas B.R.A.H.M.S. PCT™ test (PCT-V).

Design And Methods: Our laboratory previously designed a novel CLIA kit and supporting instrument (AE-180) for the detection of PCT. We analyzed the clinical performance of this system, including the imprecision, limit of detection, and linearity of analyses of 305 serum specimens. The results were compared with measurements of the same serum samples obtained with PCT-V.

Results: The limit of detection and blank of our kit were 0.0075 and 0.0039 ng/mL, respectively. The intra- and inter-assay coefficient of variation of the kit were both between 0.8% and 3.9%. The equation of linearity was found to be y = 1.03× + 0.06 (r = 0.99) for concentrations in the range of 0.01-110 ng/mL. The correlation coefficient with the results of PCT-V was 0.995, and the equation obtained for Passing and Bablok regression analysis was 1.061 for our CLIA PCT kit and - 0.003 for PCT-V. Our kit slightly overestimated the concentration according to comparison with PCT-V results.

Conclusion: The kit that was previously developed in our laboratory for the measurement of serum PCT concentration using CLIA technology shows excellent performance, just that the functional sensitivity is not as good as the PCT-V; therefore, we suggest that this kit is suitable for clinical use.

Citing Articles

Progress in Procalcitonin Detection Based on Immunoassay.

Huang J, Zu Y, Zhang L, Cui W Research (Wash D C). 2024; 7:0345.

PMID: 38711476 PMC: 11070848. DOI: 10.34133/research.0345.


Micromotor-based dual aptassay for early cost-effective diagnosis of neonatal sepsis.

Pidal J, Arruza L, Moreno-Guzman M, Lopez M, Escarpa A Mikrochim Acta. 2024; 191(2):106.

PMID: 38240873 PMC: 10798920. DOI: 10.1007/s00604-023-06134-x.


In Situ Synthesis of Highly Fluorescent, Phosphorus-Doping Carbon-Dot-Functionalized, Dendritic Silica Nanoparticles Applied for Multi-Component Lateral Flow Immunoassay.

Hu J, Ding S Sensors (Basel). 2024; 24(1).

PMID: 38202881 PMC: 10780618. DOI: 10.3390/s24010019.


A highly sensitive immunosensor based on nanochannel-confined nano-gold enhanced electrochemiluminescence for procalcitonin detection.

Chang Q, Gu X, He L, Xi F Front Chem. 2023; 11:1274424.

PMID: 37876852 PMC: 10591179. DOI: 10.3389/fchem.2023.1274424.


Procalcitonin detection in human plasma specimens using a fast version of proximity extension assay.

Bedin F, Benoit V, Ferrazzi E, Aufradet E, Boulet L, Rubens A PLoS One. 2023; 18(2):e0281157.

PMID: 36795693 PMC: 9934411. DOI: 10.1371/journal.pone.0281157.