» Articles » PMID: 32648326

Lesion Characteristics Between Cryoballoon Ablation and Radiofrequency Ablation with a Contact Force-sensing Catheter: Late-gadolinium Enhancement Magnetic Resonance Imaging Assessment

Abstract

Background: Pulmonary vein isolation (PVI) lesions after cryoballoon ablation (CBA) are characterized as a wider and more continuous than that after conventional radiofrequency catheter ablation (RFCA) without the contact force (CF)-sensing technology. However, the impact on the lesion characteristics of ablation with a CF-sensing catheter has not been well discussed. We sought to assess the lesions using late-gadolinium enhancement magnetic resonance imaging (LGE-MRI) and to compare the differences between the two groups (CB group vs. RF group).

Methods: A total of 30 consecutive patients who underwent PVI were enrolled (CB group, 18; RF group, 12). The RF applications were delivered with a target lesion size index (LSI) of 5. The PVI lesions were assessed by LGE-MRI 3 months after the PVI. The region around the PV was divided into eight segments: roof, anterior-superior, anterior carina, anterior inferior, bottom, posterior inferior, posterior carina, and posterior superior segment. The lesion width and visual gap of each segment were compared between the two groups. The visual gaps were defined as no-enhancement site of >4 mm.

Results: The mean LSI was 4.7 ± 0.7. The lesion width was significantly wider but the visual gaps were more frequently documented at the bottom segment of right PV in the CBA group (lesion width: 8.1 ± 2.2 vs. 6.3 ± 2.2 mm; p = .032; visual gap at the bottom segment or right PV: 39% vs. 0%; p = .016).

Conclusions: The PVI lesion was wider after CBA, while the visual gaps were fewer after RFCA with a CF-sensing catheter.

Citing Articles

Magnetic Resonance Imaging Monitoring of Thermal Lesions Produced by Focused Ultrasound.

Antoniou A, Evripidou N, Nikolaou A, Georgiou A, Giannakou M, Chrysanthou A J Med Ultrasound. 2025; 32(4):297-308.

PMID: 39801543 PMC: 11717076. DOI: 10.4103/jmu.jmu_112_23.


Anti- and pro-fibrillatory effects of pulmonary vein isolation gaps in human atrial fibrillation digital twins.

Jin Z, Hwang T, Kim D, Lim B, Kwon O, Kim S NPJ Digit Med. 2024; 7(1):81.

PMID: 38532181 PMC: 10966060. DOI: 10.1038/s41746-024-01075-y.


Pulmonary vein capture is a predictor for long-term success of stand-alone pulmonary vein isolation with cryoballoon ablation in patients with persistent atrial fibrillation.

Babak A, Kauffman C, Lynady C, McClellan R, Venkatachalam K, Kusumoto F Front Cardiovasc Med. 2024; 10:1150378.

PMID: 38410505 PMC: 10895012. DOI: 10.3389/fcvm.2023.1150378.


Evaluation of Pulmonary Vein Fibrosis Following Cryoballoon Ablation of Atrial Fibrillation: A Semi-Automatic MRI Analysis.

Ballatore A, Negrello E, Gatti M, Matta M, Desalvo P, Marcialis L J Cardiovasc Dev Dis. 2023; 10(9).

PMID: 37754825 PMC: 10531798. DOI: 10.3390/jcdd10090396.


Pulmonary vein isolation for atrial fibrillation using true high-power short-duration vs. cryoballoon ablation.

Wormann J, Luker J, van den Bruck J, Filipovic K, Erlhofer S, Scheurlen C Clin Res Cardiol. 2023; 112(6):846-852.

PMID: 37009942 PMC: 10241727. DOI: 10.1007/s00392-023-02188-2.