» Articles » PMID: 32476191

Digital Health Care Solution for Proactive Heart Failure Management with the Cordella Heart Failure System: Results of the SIRONA First-in-human Study

Overview
Publisher Wiley
Date 2020 Jun 2
PMID 32476191
Citations 32
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: Incorporation of remote monitoring of pulmonary artery pressure and vital signs has been demonstrated to reduce heart failure (HF) hospitalization and all-cause mortality in selected symptomatic HF patients. The aim of this study is to investigate the safety and accuracy of the new Cordella Pulmonary Artery Pressure Sensor (Endotronix, Inc., Chicago, IL, USA) and the usability of the comprehensive Cordella Heart Failure System (CHFS).

Methods And Results: Multicentre, open-label, first-in-human, feasibility study to evaluate the CHFS and the safety and accuracy of the Cordella™ Pulmonary Artery Pressure Sensor in 15 patients with New York Heart Association class III HF. All patients were successfully implanted with the Cordella Pulmonary Artery Pressure Sensor, without sensor failure. No device system-related complications, defined as invasive treatment, device explant or death, occurred. The primary efficacy endpoint of a mean pulmonary artery pressure at 90 days was met in all but one patients with a cohort difference of 2.7 mmHg (Cordella Sensor 22.5 ± 11.8 mmHg, Swan-Ganz catheter 25.2 ± 8.5 mmHg). One patient did not go through the 90-day right heart catheterization for safety reasons. Patient adherence to daily measurement, transmission of vital signs and pulmonary artery pressure sensor readings were recorded 99% of the time.

Conclusion: The initial experience of the CHFS incorporating comprehensive vital signs and pulmonary artery pressure monitoring enables safe and accurate monitoring of HF status.

Citing Articles

Clinical Characteristics and Outcomes in Heart Failure Patients with Implantable Pulmonary Artery Pressure Monitors: A Single Centre Irish Experience.

Leahy N, OBrien C, Alsubai S, Coen E, Murphy D, Sharif F J Cardiovasc Dev Dis. 2025; 12(1).

PMID: 39852303 PMC: 11766364. DOI: 10.3390/jcdd12010025.


Precision Medicine for Pulmonary Vascular Disease: The Future Is Now (2023 Grover Conference Series).

Forbes L, Bauer N, Bhadra A, Bogaard H, Choudhary G, Goss K Pulm Circ. 2025; 15(1):e70027.

PMID: 39749110 PMC: 11693987. DOI: 10.1002/pul2.70027.


Telemonitoring for Chronic Heart Failure: Narrative Review of the 20-Year Journey From Concept to Standard Care in Germany.

Spethmann S, Hindricks G, Koehler K, Stoerk S, Angermann C, Bohm M J Med Internet Res. 2024; 26:e63391.

PMID: 39631073 PMC: 11656115. DOI: 10.2196/63391.


Device therapies for heart failure with reduced ejection fraction: a new era.

Mody R, Nee Sheth A, Dash D, Mody B, Agrawal A, Monga I Front Cardiovasc Med. 2024; 11:1388232.

PMID: 39494238 PMC: 11527719. DOI: 10.3389/fcvm.2024.1388232.


Hemodynamic Monitoring Devices in the Management of Outpatient Heart Failure.

Lundgren S, Garvin R, Abraham W J Soc Cardiovasc Angiogr Interv. 2024; 2(6Part B):101186.

PMID: 39131070 PMC: 11307516. DOI: 10.1016/j.jscai.2023.101186.


References
1.
Koehler F, Koehler K, Deckwart O, Prescher S, Wegscheider K, Kirwan B . Efficacy of telemedical interventional management in patients with heart failure (TIM-HF2): a randomised, controlled, parallel-group, unmasked trial. Lancet. 2018; 392(10152):1047-1057. DOI: 10.1016/S0140-6736(18)31880-4. View

2.
Seferovic P, Ponikowski P, Anker S, Bauersachs J, Chioncel O, Cleland J . Clinical practice update on heart failure 2019: pharmacotherapy, procedures, devices and patient management. An expert consensus meeting report of the Heart Failure Association of the European Society of Cardiology. Eur J Heart Fail. 2019; 21(10):1169-1186. DOI: 10.1002/ejhf.1531. View

3.
Heywood J, Jermyn R, Shavelle D, Abraham W, Bhimaraj A, Bhatt K . Impact of Practice-Based Management of Pulmonary Artery Pressures in 2000 Patients Implanted With the CardioMEMS Sensor. Circulation. 2017; 135(16):1509-1517. DOI: 10.1161/CIRCULATIONAHA.116.026184. View

4.
Abraham W, Adamson P, Hasan A, Bourge R, Pamboukian S, Aaron M . Safety and accuracy of a wireless pulmonary artery pressure monitoring system in patients with heart failure. Am Heart J. 2011; 161(3):558-66. DOI: 10.1016/j.ahj.2010.10.041. View

5.
Abraham W, Adamson P, Bourge R, Aaron M, Costanzo M, Stevenson L . Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial. Lancet. 2011; 377(9766):658-66. DOI: 10.1016/S0140-6736(11)60101-3. View