Shared Ventilation in the Era of COVID-19: A Theoretical Consideration of the Dangers and Potential Solutions

Overview
Journal Respir Care
Publisher Mary Ann Liebert
Specialty Pulmonary Medicine
Date 2020 May 8
PMID 32376612
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The use of shared ventilation, or the simultaneous support of multiple patients connected in parallel to a single mechanical ventilator, is receiving considerable interest for addressing the severe shortage of mechanical ventilators available during the novel coronavirus pandemic (COVID-19). In this paper we highlight the potentially disastrous consequences of naïve shared ventilation, in which patients are simply connected in parallel to a ventilator without any regard to their individual ventilatory requirements. We then examine possible approaches for individualization of mechanical ventilation, using modifications to the breathing circuit that may enable tuning of individual tidal volumes and driving pressures during either volume-controlled ventilation (VCV) or pressure-controlled ventilation (PCV).

Methods: Breathing circuit modifications included a PEEP valve on each expiratory limb for both VCV and PCV, an adjustable constriction and one-way valve on the inspiratory limb for VCV, and a pressure-relief valve for peak inspiratory pressure reduction on the inspiratory limb for PCV. The ability to regulate individual tidal volumes using these breathing circuit modifications was tested both theoretically in computer simulations as well as experimentally in mechanical test lungs.

Results: In both the simulations and experimental measurements, naïve shared ventilation resulted in large imbalances across individual tidal volume delivery, dependent on imbalances across patient mechanical properties. The proposed breathing circuit modifications for shared VCV and shared PCV enabled optimization of tidal volume distributions. Individual tidal volume for one patient during shared VCV was sensitive to changes in the mechanical properties of other patients. By contrast, shared PCV enabled independent control of individual patient-received ventilation.

Conclusions: Of the shared ventilation strategies considered, shared PCV, with the inclusion of in-line pressure-relief valves in the individual inspiratory and expiratory limbs, offers the greatest degree of safety and lowest risk of catastrophic mechanical interactions between multiple patients connected to a single ventilator.

Citing Articles

Quantitative Comparison of Ventilation Parameters of Different Approaches to Ventilator Splitting and Multiplexing.

Kim D, Roy S, McBeth P, Lee J Crit Care Explor. 2024; 6(7):e1113.

PMID: 38916647 PMC: 11208113. DOI: 10.1097/CCE.0000000000001113.


Design of a flow modulation device to facilitate individualized ventilation in a shared ventilator setup.

Stiers M, Vercauteren J, Schepens T, Mergeay M, Janssen L, Hoogmartens O J Clin Monit Comput. 2024; 38(3):679-690.

PMID: 38557919 PMC: 11164813. DOI: 10.1007/s10877-024-01138-1.


Development and evaluation of a mechanical ventilator-sharing system.

Achanta S, Gentile M, Euliano N Front Med (Lausanne). 2024; 11:1356769.

PMID: 38435386 PMC: 10905385. DOI: 10.3389/fmed.2024.1356769.


Simulated ventilation of two patients with a single ventilator in a pandemic setting.

Schepat P, Kober B, Eble M, Wenzel V, Herff H Med Gas Res. 2023; 14(2):61-66.

PMID: 37929509 PMC: 10715326. DOI: 10.4103/2045-9912.385438.


COVID-19 Lessons Learned: Response to the Anticipated Ventilator Shortage.

Branson R, Rodriquez Jr D Respir Care. 2022; 68(1):129-150.

PMID: 36566030 PMC: 9993519. DOI: 10.4187/respcare.10676.