» Articles » PMID: 32206335

The Activ8 Activity Monitor: Validation of Posture and Movement Classification

Overview
Date 2020 Mar 25
PMID 32206335
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: To set up and guide interventions with the aim to increase physical activity and lower sedentary behavior valid monitoring of physical behavior is essential. The aim of this study was to evaluate the validity of the single-unit Activ8 activity monitor to classify several body postures and movements.

Methods: Twelve healthy adults performed a series of activities, representative for everyday life, according to a standard protocol. Activ8 was both worn in the trouser pocket (prescribed location) and fixated to the front of the thigh. Activities were video recorded and analyzed thereafter. Postures and movements that were analyzed were lying/sitting, standing, walking, cycling, and running.

Results: The agreement between Activ8 output and video analysis was 89.7% (inter-subject range: 66.0 to 96.6%) for the pocket location and 91.9% (range 85.5 to 95.1%) for the thigh location. Sensitivity and positive predictive value scores for both locations were all above 80%, except for standing (69% or higher). Differences in classified duration of separate postures and movements were within 20% for walking, sitting and running.

Conclusion: The Activ8 is a valid instrument to quantify a defined set of body postures and movements. Because of the smaller time difference, the thigh location is preferred for research purposes.

Citing Articles

Effect of a soft exosuit on daily life gait performance in people with incomplete spinal cord injury: study protocol for a randomized controlled trial.

Visch L, Groen B, Geurts A, van Nes I, Keijsers N Trials. 2024; 25(1):592.

PMID: 39242508 PMC: 11378477. DOI: 10.1186/s13063-024-08412-2.


Correlates of physical activity in ambulatory people with spinal cord injury during the first year after inpatient rehabilitation.

Postma K, van Diemen T, Post M, Stolwijk-Swuste J, van den Berg-Emons R, Osterthun R Spinal Cord. 2024; 62(5):249-254.

PMID: 38509176 DOI: 10.1038/s41393-024-00982-x.


Validation of the Activ8 Activity Monitor for Monitoring Postures, Motions, Transfers, and Steps of Hospitalized Patients.

Becker M, Hurkmans H, Verhaar J, Bussmann J Sensors (Basel). 2024; 24(1).

PMID: 38203041 PMC: 10781347. DOI: 10.3390/s24010180.


Effectiveness and Cost-Effectiveness of a Stratified Blended Physiotherapy Intervention Compared With Face-to-Face Physiotherapy in Patients With Nonspecific Low Back Pain: Cluster Randomized Controlled Trial.

Koppenaal T, Van Dongen J, Kloek C, Arensman R, Veenhof C, Pisters M J Med Internet Res. 2023; 25:e43034.

PMID: 37999947 PMC: 10709796. DOI: 10.2196/43034.


A DEDICATED AMPUTEE SPORTS PROGRAMME IMPROVES PHYSICAL FUNCTIONING AND SPORTS PARTICIPATION.

VAN Oorschot W, VAN Ee I, Keijsers N J Rehabil Med Clin Commun. 2023; 6:12392.

PMID: 37964806 PMC: 10642343. DOI: 10.2340/jrmcc.v6.12392.


References
1.
Reilly J, Penpraze V, Hislop J, Davies G, Grant S, Paton J . Objective measurement of physical activity and sedentary behaviour: review with new data. Arch Dis Child. 2008; 93(7):614-9. DOI: 10.1136/adc.2007.133272. View

2.
Bassett Jr D, Rowlands A, Trost S . Calibration and validation of wearable monitors. Med Sci Sports Exerc. 2011; 44(1 Suppl 1):S32-8. PMC: 3273335. DOI: 10.1249/MSS.0b013e3182399cf7. View

3.
Russchen H, Slaman J, Stam H, Markus-Doornbosch F, van den Berg-Emons R, Roebroeck M . Focus on fatigue amongst young adults with spastic cerebral palsy. J Neuroeng Rehabil. 2014; 11:161. PMC: 4274699. DOI: 10.1186/1743-0003-11-161. View

4.
Pavey T, Gomersall S, Clark B, Brown W . The validity of the GENEActiv wrist-worn accelerometer for measuring adult sedentary time in free living. J Sci Med Sport. 2015; 19(5):395-9. DOI: 10.1016/j.jsams.2015.04.007. View

5.