Flow Rate Variance of a Fully Implantable Pump for the Delivery of Intravenous Treprostinil in Pulmonary Arterial Hypertension
Overview
Authors
Affiliations
Implantable infusion pumps might improve the convenience and safety of intravenous treprostinil for pulmonary arterial hypertension. The LENUS Pro® pump (approved in Europe) has a fixed flow rate. Based on 126 pumps and 2853 refills, we retrospectively analyzed the actual flow rate from 09/2010 to 09/2018. A relevant flow rate variance is evident after three years; therefore, flow rate monitoring and dose adjustment are mandatory.
Kamp J, Fuge J, Karsten J, Rumke S, Hoeper M, Park D BMC Pulm Med. 2021; 21(1):164.
PMID: 33992098 PMC: 8126130. DOI: 10.1186/s12890-021-01541-3.
Steringer-Mascherbauer R, Lummersdorfer M, Fugger R, Sigmund E, Huber C, Engleder D Pulm Circ. 2020; 10(4):2045894020970830.
PMID: 33299542 PMC: 7711148. DOI: 10.1177/2045894020970830.
Kopp G, Hachulla A, Noble S, Bringard A, Soccal P, Beghetti M Front Med (Lausanne). 2020; 7:539707.
PMID: 33195292 PMC: 7662390. DOI: 10.3389/fmed.2020.539707.