» Articles » PMID: 32015737

Biplanar Low-Dose Radiography Is Accurate for Measuring Combined Anteversion After Total Hip Arthroplasty

Overview
Journal HSS J
Date 2020 Feb 5
PMID 32015737
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Acetabular component position alone has not been predictive of stability after total hip arthroplasty (THA). Combined anteversion of the acetabulum and femur has the potential of being more predictive of stability. Unfortunately, femoral component position is difficult to measure on plain radiographs. Computed tomography (CT) is the gold standard for measuring implant position post-operatively, but CT exposes patients to a substantial amount of radiation.

Questions/purposes: We sought to determine whether biplanar low-dose radiography can be used to accurately measure both acetabular and femoral implant position after THA.

Methods: Twenty patients underwent standing low-dose biplanar spine-to-ankle radiographs and supine CT scans 6 weeks after THA. Measurements of acetabular inclination, acetabular anteversion, and femoral anteversion were performed by two blinded observers and compared.

Results: The average absolute differences between biplanar radiographs and CT scans were 2° ± 2° for acetabular inclination, 3° ± 2° for acetabular anteversion, and 4° ± 4° for femoral anteversion between EOS measurements and CT measurements. Interobserver agreement was good for acetabular inclination, acetabular anteversion, and femoral anteversion (Cronbach's  = 0.90) using biplanar low-dose imaging.

Conclusion: Biplanar radiography is a reliable low-radiation alternative for measuring acetabular inclination, acetabular anteversion, femoral version, and thus combined anteversion compared to CT. Femoral anteversion had the most variability but is still clinically relevant.

Citing Articles

"The accuracy of the EOS imaging system to assess hip abnormalities in adolescents and adults:" a systematic review and meta-analysis.

Alghamdi A, Madan S, Shuweihdi F, Offiah A Skeletal Radiol. 2023; 53(1):29-42.

PMID: 37338589 PMC: 10661804. DOI: 10.1007/s00256-023-04351-2.


Reliability and validity of the functional combined anteversion measurement method using standing lateral radiography after total hip arthroplasty.

Zhang W, Li D, Xu J, Sun H, Cai Z, Chen M Ann Transl Med. 2023; 11(5):196.

PMID: 37007575 PMC: 10061452. DOI: 10.21037/atm-22-3243.


The role of the femoral component orientation on dislocations in THA: a systematic review.

van Erp J, Snijders T, Weinans H, Castelein R, Schlosser T, de Gast A Arch Orthop Trauma Surg. 2021; 142(6):1253-1264.

PMID: 34101017 PMC: 9110501. DOI: 10.1007/s00402-021-03982-1.


Validity of the EOS-determined pelvic parameters and orientation with pelvic positional variation: a phantom study.

Kim J, Lee D, Lee H, Shin H, Park B, Park S Sci Rep. 2021; 11(1):10468.

PMID: 34001983 PMC: 8129068. DOI: 10.1038/s41598-021-89958-y.


A review of plain radiographic methods for measuring femoral version after total hip arthroplasty and validation of modified Budin's method in Indian population.

Mittal A, Kumar N, Nandimandalam B, Pankaj A J Clin Orthop Trauma. 2021; 15:104-109.

PMID: 33717923 PMC: 7920004. DOI: 10.1016/j.jcot.2020.10.019.


References
1.
Tsai T, Dimitriou D, Li G, Kwon Y . Does total hip arthroplasty restore native hip anatomy? three-dimensional reconstruction analysis. Int Orthop. 2014; 38(8):1577-83. PMC: 4115107. DOI: 10.1007/s00264-014-2401-3. View

2.
Lin F, Lim D, Wixson R, Milos S, Hendrix R, Makhsous M . Validation of a computer navigation system and a CT method for determination of the orientation of implanted acetabular cup in total hip arthroplasty: a cadaver study. Clin Biomech (Bristol). 2008; 23(8):1004-11. DOI: 10.1016/j.clinbiomech.2008.04.015. View

3.
McArthur B, Cross M, Geatrakas C, Mayman D, Ghelman B . Measuring acetabular component version after THA: CT or plain radiograph?. Clin Orthop Relat Res. 2012; 470(10):2810-8. PMC: 3441984. DOI: 10.1007/s11999-012-2292-9. View

4.
Guenoun B, El Hajj F, Biau D, Anract P, Courpied J . Reliability of a new method for evaluating femoral stem positioning after total hip arthroplasty based on stereoradiographic 3D reconstruction. J Arthroplasty. 2014; 30(1):141-4. DOI: 10.1016/j.arth.2014.07.033. View

5.
Lewinnek G, Lewis J, Tarr R, COMPERE C, Zimmerman J . Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978; 60(2):217-20. View